5. OPERATION PROCEDURES

 

 
Attaching air turbine handpiece:   
 
1. Remove the bur from the handpiece.   
2. Open the door of the unit 
3. Attach the handpiece to the joint and tighten the nut securely 
4. Close the door securely. 
 

CAUTION 

Be sure to remove the bur from the handpiece before connecting it to each joint., 
 

 

 

Straight Handpiece 
 
1. For the straight handpiece, attach the supplied test bur and lock it 
2. Open the door of the unit and attach the handpiece to the E-type joint until it locks 
3. Close the door securely. 
 
Contra Angle Handpiece

 

 
1. Remove the bur from the handpiece. 
2. Open the door of the unit and attach the handpiece to the E-type joint until it locks 
3. Point the bur insertion hole toward the door side and close the door securely. 

 

CAUTION 

 

Be sure to remove the bur from the handpiece before connecting to each joint. For the straight 
handpiece, be sure to attach the supplied test bur and lock it before connecting the handpiece to 
each joint. 

 

For the contra angle handpiece, point the bur insertion hole toward the door side and close the 
door. Not pointing the bur insertion hole toward the door side may not allow the door to be 
closed with contact during closing

 

 

 

Содержание HCR-100

Страница 1: ...1 iM3 HCR 100 Automatic Handpiece Cleaning Lubrication System OPERATION MANUAL...

Страница 2: ...UTIONS 3 1 FEATURES 5 2 SPECIFICATIONS 6 3 PARTS NAME 7 4 INSTALLATION AND ASSEMBLY 7 5 OPERATION PROCEDURES 8 6 DOOR 11 7 MAINTENANCE 11 8 TROUBLE SHOOTIING 12 SAFETY PRECAUTIONS Read these safety ca...

Страница 3: ...use the product in direct sunlight or where the temperature exceeds 40C 104F Do not use the product for other purposes than the maintenance cleaning and lubrication of handpieces described in this op...

Страница 4: ...sure above 0 35Mpa Do not open the oil cap during operation Open only after the unit completely stops When opening the oil cap in case of abnormal stop opening the door wait for approx 5 seconds befor...

Страница 5: ...r effectively and efficiently cleans and lubricates contra angle handpiece Pressing the air key can purge excess oil from the handpiece after lubrication and cleaning cycle to some extent Select the l...

Страница 6: ...6 2 SPECIFICATIONS Model SC L1 Rated Voltage AC120V10 AC230V10 50 60Hz Input Voltage 24 0VA Air Pressure 0 25 0 3Mpa 2 5 3 0kgf cm2 50 85psi Dimensions W280 D275 H360mm Weight 6 8kg...

Страница 7: ...ll it to the full level while checking the oil level gauge After filling oil screw the oil cap back on tight DANGER Do not fill maintenance oil in the room with a risk of an explosion or near open fla...

Страница 8: ...lose the door securely Contra Angle Handpiece 1 Remove the bur from the handpiece 2 Open the door of the unit and attach the handpiece to the E type joint until it locks 3 Point the bur insertion hole...

Страница 9: ...cation cycle 5 When the cleaning and lubrication cycle is completed the mode indicator turns on and the buzzer sounds to tell the completion of the cycle 6 Pressing the air key blows out air only from...

Страница 10: ...ng the air key 2 In 1 2 seconds or less press the start key The mode indicator changes at that time S L lighting The buzzer sounds L E lighting The buzzer does not sound 3 After selecting either press...

Страница 11: ...leaning cycle 4 or 5 times before use to secure stable oil absorption of mist filters Alternatively apply maintenance oil directly to mist filters before use CAUTION Replace oil absorbent sheets appro...

Страница 12: ...burned The unit does not clean and lubricate Does the mode indicator light Is the door open Remove the air tube press all three selection buttons and press start button Does the Unit produce an operat...