5. OPERATION PROCEDURES
Attaching air turbine handpiece:
1. Remove the bur from the handpiece.
2. Open the door of the unit
3. Attach the handpiece to the joint and tighten the nut securely
4. Close the door securely.
CAUTION
Be sure to remove the bur from the handpiece before connecting it to each joint.,
Straight Handpiece
1. For the straight handpiece, attach the supplied test bur and lock it
2. Open the door of the unit and attach the handpiece to the E-type joint until it locks
3. Close the door securely.
Contra Angle Handpiece
1. Remove the bur from the handpiece.
2. Open the door of the unit and attach the handpiece to the E-type joint until it locks
3. Point the bur insertion hole toward the door side and close the door securely.
CAUTION
Be sure to remove the bur from the handpiece before connecting to each joint. For the straight
handpiece, be sure to attach the supplied test bur and lock it before connecting the handpiece to
each joint.
For the contra angle handpiece, point the bur insertion hole toward the door side and close the
door. Not pointing the bur insertion hole toward the door side may not allow the door to be
closed with contact during closing