CAUTION
Always autoclave handpieces after they have been cleaned and lubricated with this product.
Ideal operating environment is 10-40 C (50-104 F), humidity at 25-70%RH, atmospheric pressure at
500-1060hPa and no moisture condensation in the Unit. Use outside of these conditions may cause
malfunction.
Ideal storage environment is 10-50 C (14-122 f), humidity at 10-85%RH, atmospheric pressure at
500-1060hPa and the system is not subject to air with dust, sulfur or salinity.
When installing the product leave a space of approximately 10cm around the unit.
Place a collector such as a tray under the unit.
Set the air pressure between 0.25-0.30Mpa
Do not set the air pressure above 0.35Mpa
Do not open the oil cap during operation. Open only after the unit completely stops.
When opening the oil cap in case of abnormal stop (opening the door), wait for approx. 5 seconds
before opening cap.
After use, immediately turn off the power switch and close the air supply.
The air supply to this device (air compressor etc.) and air filter need to be clean and free from
moisture. Drain the air compressor and air filter at least once a week. Moisture from the tank could
mix in the tank could mix in the maintenance oil and defeat the purpose of this device.
Remove the air tube and the power cord if the unit will not be used for a long time.
When storing the maintenance oil, keep away from strong oxidizing agents, combustible materials
and strong acid chemicals.
Only purchase replacement parts and consumables from iM3 or risk invalidating the warranty.