I N S T A L L A T I O N A N D O P E R A T I O N
Front Panel Controls
16
OMM-6810B
C H A P T E R
2
Wavelength Display Units
Delta Power Display Control
When the DISPLAY
P button is pressed, the power at that instant becomes the
reference and the left display begins indicating the change in power from that
reference.
P can be displayed in linear units (
P and the appropriate linear units
indicator are illuminated) or logarithmic units (
P and the dB indicators are
illuminated). A negative change in power indicates a power lower than the
reference power (pressing the DISPLAY
P button again returns the power display
to absolute measurements).
Delta Lambda Display Control
When the DISPLAY
Δ
button is pressed, the wavelength measured at that instant
becomes the reference and the display begins indicating the change in
wavelength from that reference.
Δ
may be displayed in nanometers (
Δ
and nm
indicators are illuminated) or wavenumbers (
Δ
and cm
-1
indicators are
illuminated). Pressing the DISPLAY
Δ
button again returns the wavelength
display to absolute measurements. A negative change in wavelength indicates a
shorter wavelength with respect to the reference.
The DISPLAY nm/cm
-1
button toggles the wavelength
display between nanometers (nm) and wavenumbers (cm
-1
).
The DISPLAY
P button is used to set a reference power
and begin displaying the change in power from that
reference.
The DISPLAY
Δ
button is used to set a reference
wavelength and begin displaying the change in wavelength
from that reference.