C O M M A N D R E F E R E N C E
Command Paths
01_10
OMM-6810B
95
C H A P T E R
4
The WAVE:REFerence command is used to set the wavelength reference point for future
delta wavelength measurements.
Parameters
The <nrf> parameter is a floating point number representing the reference wavelength.
The <suffix> is an approved wavelength suffix as shown in Table 3.2. If the suffix is
omitted, nanometers is assumed.
Notes
The display mode should be in DISPLAY
mode before this command is sent. Sending
the WAVE:MODE DELTA ON command or pressing the DISPLAY
button assigns a
reference wavelength. Use this command to change the reference point while in the
mode.
The wavelength reference point may be queried by sending WAVE:REFerence?
Examples
“WAVE:REFERENCE 15000 CM” - action: the reference wavelength is set to
15000 cm
-1
.
“WAVE:REF 632.8” - action: the reference wavelength is set to 632.8 nm.
WAVE:REFerence
F
RONT
P
ANEL
R
EMOTE
:
<white
space>
<nrf>
:
WAVE
REF
<white
space>
<suffix>