I N T R O D U C T I O N A N D S P E C I F I C A T I O N S
Specifications: LDX-3545
3_03
LDX-3500 Series
5
C H A P T E R
1
Specifications: LDX-3545
1. Stability specifications are measured at half-scale output, after a one hour warm-up period.
2. Extrapolated from resulting intensity fluctuations of a laser diode, measured optically with a 150 kHz bandwidth photodetector.
3. The instrument contains a circuit which will turn off the output if a transient is detected.
4. Maximum output current transients resulting from normal operational situations (e.g., power on-off) as well as accidental situations
(e.g., power line plug removal)
5. Maximum output current transients resulting from a 200 V power-line spike.
6. Specified values are a percent of nominal. Constant-power mode stability specification assumes zero drift in detector responsivity.
Current Source
1000 mA Range
3000 mA Range
Set Point Accuracy
+1 mA
+3 mA
Set Point Resolution
250
µ
A
750
µ
A
Compliance Voltage (fixed)
6.5 V maximum
6.5 V maximum
Temperature Coefficient
< 50 ppm/
o
C
< 50 ppm/
o
C
Stability
1
, for 1 hour
< 20 ppm
< 20 ppm
Stability
1
, for 24 hours
< 50 ppm
< 50 ppm
Noise and Ripple
2
High Bandwidth Mode
CW Mode
< 8
µ
A rms
< 8
µ
A rms
< 20
µ
A rms
< 15
µ
A rms
Worst Case Transients
3
Operational
4
Power-line induced
5
< 2 mA
< 10 mA
< 5 mA
< 10 mA
Photodiode Feedback
Range
5 to 9999
µ
A
Output Stability
6
+0.02%
Set Point Accuracy
+10
µ
A
Bias Voltage
0 - 5 V reverse bias (+ 10%); adjustable on back panel
Power Modulation Bandwidth
1 kHz
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com