The TE_OPEN? query returns the current status of the sensor open error.
Parameters
None.
Notes
This error will disable the output. The status of this error is reset when the output is
enabled.
Examples
"TE_OPEN?” - response: 0, means that the TE is connected correctly and the output is
enabled.
“TE_OPEN?” - response: 1, means that the TE is not connected or open circuited.
The TEC:C1 command sets the TECs Steinhart-Hart equation constant C1.
The TEC:C1? query returns the value of the Steinhart-Hart equation constant C1.
Parameters
Steinhart-Hart value between the value of 9.999 and -9.99.
Notes
When the LM335 or AD590 sensors are selected via the SENSOR SELECT switch, only
C1 and C2 are used.
Examples
"TEC:C1 1.125” - action: sets the Steinhart-Hart constant C1 for the current selected
sensor to 1.125.
“TEC:C1?” - response: 2.100, means the Steinhart-Hart constant C1 is set to 2.100.
The TEC:C2 command sets the TECs Steinhart-Hart equation constant C2.
The TEC:C2? query returns the value of the Steinhart-Hart equation constant C2.
Parameters
Steinhart-Hart value between the value of 9.999 and -9.99.
Notes
When the LM335 or AD590 sensors are selected via the SENSOR SELECT switch, only
C1 and C2 are used.
Examples
"TEC:C2 1.125” - action: sets the Steinhart-Hart constant C2 for the current selected
sensor to 1.125.
“TEC:C2?” - response: 2.100, means the Steinhart-Hart constant C2 is set to 2.100.
TE_OPEN?
F
RONT
P
ANEL
R
EMOTE
TEC:C1
TEC:C1?
F
RONT
P
ANEL
R
EMOTE
TEC:C2
TEC:C2?
F
RONT
P
ANEL
R
EMOTE
Содержание LDT-5500B Series
Страница 2: ......
Страница 6: ...TA B L E O F C O N T E N T S iv LDT 5500B Series...
Страница 8: ...L I S T O F FIG U R E S vi LDT 5500B Series...
Страница 20: ...I N T R O D U C T I O N A N D S P E C I F I C A T I O N S Specifications 6 LDT 5500B Series C H A P T E R 1...
Страница 34: ...RE M O T E O PE R A T I ON Error Messages 20 LDT 5500B Series C H A P T E R 3...
Страница 58: ...M A I N T E N A NC E A N D TR O U B L E S H O O T I N G Troubleshooting 44 LDT 5500B Series C H A P T E R 5...
Страница 60: ...46 LDT 5500B Series C H A P T E R A...
Страница 70: ...56 LDT 5500B Series C H A P T E R C...