R E M O T E O P E R A T I O N
Command Syntax
18
LDT-5500B Series
C H A P T E R
3
Once installation is complete, plug the instrument into the computer with the USB
cable. If the Found New Hardware Wizard appears, select “Automatically Install
Recommended Software”, click NEXT, and wait for the installation to complete.
Command Syntax
This section details the syntax of the commands as expected by the LDT-5500B.
Letters
Any remote command or query must contain all of the letters of the command. The
LDT-5500B does not distinguish between capital and lower case letters.
White Space
Spaces or white space may be used to separate data but may not be used as a
separation between command and query question mark. Spaces or white space must
be used to separate the command from the first parameter.
Command Termination
Each USB command or query must always be terminated with the new line character:
'\n' or 0x0A.
Boolean Parameter Values
Boolean values (0 and 1) are used to represent On/Off or True/False.
Command Timing and Completion
All commands are executed in a sequential manner.
Содержание LDT-5500B Series
Страница 2: ......
Страница 6: ...TA B L E O F C O N T E N T S iv LDT 5500B Series...
Страница 8: ...L I S T O F FIG U R E S vi LDT 5500B Series...
Страница 20: ...I N T R O D U C T I O N A N D S P E C I F I C A T I O N S Specifications 6 LDT 5500B Series C H A P T E R 1...
Страница 34: ...RE M O T E O PE R A T I ON Error Messages 20 LDT 5500B Series C H A P T E R 3...
Страница 58: ...M A I N T E N A NC E A N D TR O U B L E S H O O T I N G Troubleshooting 44 LDT 5500B Series C H A P T E R 5...
Страница 60: ...46 LDT 5500B Series C H A P T E R A...
Страница 70: ...56 LDT 5500B Series C H A P T E R C...