C A L I B R A T I O N A N D T R O U B L E S H O O T I N G
Troubleshooting Guide
80
LDC-3926339
C H A P T E R
5
Remote Accessible Error Codes
This section contains error codes that the module reports in response to the
"MODERR?" query. Refer to the LDC-3926 Manual, for a list of mainframe error
codes and descriptions (the codes returned in response to the ERR? query).
Testing for Errors in Remote Operation
For more information about specific GPIB commands, see Chapter 4,
Command
Reference
.
1
Send the query
ERR?
to read the system errors and module error summary.
This allows you to error-check the LDC-3926 as a whole. If any module errors are
present, the corresponding bit of the module error summary is set. For example, suppose
the mainframe responds to an ERR? query with the string 0,0000000001100000. The
zero to the left of the comma indicates that there are no mainframe errors, and the binary
representation to the right of the comma indicates that there are errors on channels 7 and
6. (Module 16 is on the left, module 1 is on the right).
2
Send the query
MODERR?
to read the module errors.
For example, type
CHAN 7;MODERR?
to return the errors in module 7, and
CHAN 6;MODERR?
returns any errors in module 6. For more information about
mainframe errors, refer to the LDC-3926 Manual.
Table 5.7
Error Codes
Error Code
Explanation
E-103
Length of arbitrary block is different from expected length.
E-104
Parameter is an undefined numeric type.
E-105
Parameter has an invalid exponent.
E-106
A digit was expected in the parameter but was not found.
E-114
Specified arbitrary block length is invalid.
E-123
Command is not found.
E-126
Wrong number of parameters for command.
E-201
Parameter value out of range.
E-202
Error in conversion of parameter type.
E-203
Command is a "secure" command, but secure commands are disabled.
E-204
Suffix is invalid.
E-205
Expected Boolean parameter is invalid.
E-206
Error in conversion to signed 16-bit integer.
E-207
Error in conversion to unsigned 16-bit integer.
E-208
Error in conversion to signed 32-bit integer.
E-209
Error in conversion to unsigned 32-bit integer.
E-210
Error in conversion to floating-point number.
E-211
Error in conversion to character pointer.
E-212
Error in conversion to byte pointer.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com