R E M O T E O P E R A T I O N S
Error Messages
02_03
LDC-3926339
35
C H A P T E R
3
Error Messages
This section contains descriptions of the errors that are specific to the LDC-
3926339 module. These are the error codes that are returned via the
"MODERR?" query. Refer to the mainframe manual for a list of mainframe error
codes and descriptions (the codes returned via the "ERR?" query).
Testing for Errors in Remote Operation
For more information about specific GPIB commands, see Chapter 4,
Command
Reference
.
1
Send the query
ERR?
to read the system errors and module error summary.
This allows you to error-check the LDC-3926 as a whole. If any module errors are
present, the corresponding bit of the module error summary is set. For example, suppose
the mainframe responds to an ERR? query with the string 0,0000000001100000. The
zero to the left of the comma indicates that there are no mainframe errors, and the binary
representation to the right of the comma indicates that there are errors on channels 7 and
6.
2
Send the query
MODERR?
to read the module errors.
For example, type
CHAN 7;MODERR?
to return the errors in module 7, and
CHAN
6;MODERR?
returns any errors in module 6. For more information about mainframe
errors, refer to the LDC-3926 Manual.
Note:
Error codes not listed are reserved for future design use.
Table 3.4
Error Message Codes
Error Code
Explanation
E-103
Length of arbitrary block is different from expected length.
E-104
Parameter is an undefined numeric type.
E-105
Parameter has an invalid exponent.
E-106
A digit was expected in the parameter but was not found.
E-114
Specified arbitrary block length is invalid.
E-123
Command is not found.
E-126
Wrong number of parameters for command.
E-201
Parameter value out of range.
E-202
Error in conversion of parameter type.
E-203
Command is a "secure" command, but secure commands are disabled.
E-204
Suffix is invalid.
E-205
Expected Boolean parameter is invalid.
E-206
Error in conversion to signed 16-bit integer.
E-207
Error in conversion to unsigned 16-bit integer.
E-208
Error in conversion to signed 32-bit integer.
E-209
Error in conversion to unsigned 32-bit integer.
E-210
Error in conversion to floating-point number.