SWIFT K11-EN, Rev.A
77
I’d like to set pause.
Refer to [Option menu].
I’d like to indicate cutting angle, optical fiber’s angle and core/clad deviation.
Refer to [Option menu].
Estimated splice loss and measured splice loss does not match.
The estimated splice loss is a calculated value so it should be used only as a
reference.
9.3 Sleeve heater
Optical fiber protecting sleeve is not contracted completely.
Increase the heating time. Refer to [Heater mode edition].
Heater is overheated.
Stop the heater by pressing
, turn the power off and then contact to
Ilsintech.
If the protecting sleeve melts and it sticks to the heater cover, remove it by
pushing with cotton swab.
I’d like to initialize heater mode’s heating condition.
Refer to [Heater mode edition].
I’d like to cancel heater in the middle of operation.
Heater operation cannot be canceled
by pressing
. Cancel it by pressing
once again.