IDS-1100 Dermatoscope
Guidance and manufacturer declaration – Electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - Guideline
Electrostatic discharge (ESD)
Acc. To IEC 61000-4-2
±6kV contact discharge
±8kV air discharge
±6kV contact discharge
±8kV air discharge
Floors should be wood, concrete or covered with
ceramic tiles. If floors are covered with synthetic
material, the relative humidity should be at least
30%
Electrical fast transient/burst
IEC 61000-4-4
not applicable
not applicable
not applicable
Surge IEC 61000-4-5
not applicable
not applicable
not applicable
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
not applicable
not applicable
not applicable
Power frequency (50/60 Hz)
Magnetic field IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
Comments: UT is the a.c. supply voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that
it is used in such environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment Guidelines
Conducted RF
not applicable
not applicable
not applicable
14