REVEXmax - User’s Manual
6
Instrument main menu
6.5
Menu - Setup
6.5.1
Menu - System info
These items can be displayed Serial number, Firmware version, Production date and
The number of records.
6.5.2
Menu - Self-test
This option can be performed to verify the functionality of the REVEXmax. Dis-
connect equipment under test before the self-test. The self-test does not replace
calibration. The procedure is described in chapter 3.2.
6.5.3
Menu - Date of next calibration
In this menu, you can set the next calibration date. The measuring device starts
to display a warning message 45 days before the expiration date, e.g.,
Date of the
next calibration: 01.01.2020
. This information is displayed only once a day. If
the calibration has expired, there is the message, e.g.,
Calibration expired on
01.01.2020
. Enter into the menu, the date of the next calibration is performed by
pressing the
[ent]
key. The date is inserted using the keyboard device. Confirm the
selection by pressing the
[ent]
key or press the
[esc]
key to return without saving
the changes.
The ’.’ characters are automatically skipped in the editor.
6.5.4
Menu - Erase database
Use this option to erase the entire main memory device, i.e., database appliances.
The essential instrument settings and a list of test sequences remain preserved.
Erasing the memory must be confirmed by pressing the
[Yes]
key.
How long it takes to erasing entries depends on the memory usage. It
can take up to 2 minutes when all memory locations are occupied.
6.5.5
Menu - Measurement locking
This option allows you to automatically lock every measurement. When enabled,
each time measurement is displayed the symbol
AUT.ARET
. The switch activation
is done by pressing the
[ent]
key.
6.5.6
Menu - Save last method
This option allows you to save last measuring method. When enabled, last method
is selected immediately after connection the REVEXmax to the mains. The switch
activation is done by pressing the
[ent]
key.