7.2 Ion Balance / Performance Check 
For checking balance and performance a charge plate monitor (CPM) such as SIMCO’S 
EA-4 is required. The CPM should have its ground connected to the instrument ground 
provided on the rear panel of the Centurion; this jack accepts a standard banana plug. 
With the charge plate located 12” from the outlet grille of the unit, the ion balance 
control can be checked. Note that if an adjustment to the Ion Balance is required, the 
emitters should be cleaned (see 7.3 below) before this adjustment takes place. By using 
the CPM to measure discharge times of the charge plate, ion content of the air and ion 
delivery to the target area can be verified.  
 
7.3 Emitter Cleaning  
Recommended frequency of cleaning is once per week. To clean ion emitters, simply 
rotate the emitter cleaner knob located at the center of the fan grille on the rear of the 
unit. Rotate the knob clockwise until it stops and release. The spring-loaded emitter 
cleaning brush will return to its parking spot.  
 
It is not necessary to turn off the unit during cleaning of the emitters when using the 
integrated cleaning brush. 

 

 

7.4 Emitter Replacement 
To replace emitters, the rear of the unit must be removed. Loosen the #8 Phillips-head 
screws located on the bottom of the chassis. The rear cover can now swing open for 
removal from its pivot pins. The emitters are installed in sockets for easy insertion and 
removal. Using a long-nose pliers, pull the emitters out along the radial direction, 
gripping the shaft only, taking care not to damage the sharp tip. Position each emitter 
carefully during insertion, as excessive force can damage the emitter sockets. Using 
long-nose pliers, push the emitters into the sockets along the radial direction.  
 
7.5 Troubleshooting / Diagnostics 

 

This information provides a quick troubleshooting reference for the Centurion single-fan 
ionizer. Should any of these possible solutions not solve the problem, contact SIMCO. 
 
 

PROBLEM CAUSE  SOLUTION 

UNIT FAILS TO OPERATE 

POWER CORD NOT CONNECTED  

PLUG IN TO SPECIFIED VOLTAGE SOURCE 

UNIT FAILS TO OPERATE 

FAULTY AC ADAPTER 

REPLACE AC ADAPTER 

FLASHING FAULT LIGHT 

POWER INCORRECTLY SUPPLIED TO 
UNIT 

CYCLE POWER TO THE UNIT USING THE 
UNIT’S REAR PANEL SWITCH 

FLASHING FAULT LIGHT 

POWER INCORRECTLY SUPPLIED TO 
UNIT 

ENSURE THAT PROPERLY GROUNDED 
RECEPTICLE IS USED 

FAULT LIGHT ON STEADY 

DIRTY OR DAMAGED EMITTERS, 
INCORRECTLY CALIBRATED 

CLEAN EMITTERS USING INTEGRATED 
BRUSH, REPLACE AS NECESSARY, 
RECALIBRATE 

EXCESSIVELY LONG DISCHARGE 
TIMES 

DIRTY OR DAMAGED EMITTERS, 
INCORRECTLY CALIBRATED 

CLEAN EMITTERS USING INTEGRATED 
BRUSH, REPLACE AS NECESSARY, 
RECALIBRATE 

UNIT FAILS TO HOLD SETPOINT 

DIRTY OR DAMAGED EMITTERS 

CLEAN EMITTERS USING INTEGRATED 
BRUSH, REPLACE AS NECESSARY, 
RECALIBRATE 

FAN IS SLOW 

FAN SPEED IS SET SLOW 

ADJUST FAN CONTROL CLOCKWISE 

 

 

Содержание SIMCO CENTURION

Страница 1: ...single fan DC ionizer offers superior balance performance with its novel emitter array design and high gain dual feedback circuitry The Centurion eliminates static charges cleanly quickly and reliably...

Страница 2: ...made to the output of the emitter array are seen promptly as ion output changes in the air The resulting balanced stream of ionized air is delivered to the work area to provide static neutralization...

Страница 3: ...an Speed Adjusted using a potentiometer adjustment tool Adjustment located on the rear panel of the unit Offset Balance Adjustable to zero using a potentiometer adjustment tool Adjustment access is lo...

Страница 4: ...e unit in flammable volatile or explosive atmosphere 6 Do not insert objects through the unit s intake or outlet grilles Damage to the ionizer and or personal injury may result SECTION 5 Installation...

Страница 5: ...ize static is dependent on the fan speed setting For fastest charge neutralizing performance the unit should be adjusted for the highest fan speed The Ion Balance adjustment located on the rear of the...

Страница 6: ...removal from its pivot pins The emitters are installed in sockets for easy insertion and removal Using a long nose pliers pull the emitters out along the radial direction gripping the shaft only takin...

Страница 7: ...n no event shall SIMCO be liable for incidental or consequential damages except where state laws override This warranty extends to the original purchaser and is not transferable No person agent distri...

Отзывы: