SECTION 4 Safety
CAUTION!
1. Read the complete Operation Manual before operating. Failure to follow instructions
may result in damage to the ionizer and/or personal injury.
2. This unit is supplied with a 3-prong grounding plug, which must be inserted in an
appropriate, properly wired, and grounded receptacle. The 24VDC AC adapter used
with the unit provides a ground connection at its output. Do not defeat the electrical
ground, grounding and proper wiring are required for proper operation.
3. Do not ground or make electrical connections of any kind to the unit’s metal fan
guards located on the intake and output.
4. A factory-qualified service technician must perform component service and repairs.
Please contact SIMCO Customer Service for information.
WARNING!
5. Keep the unit dry. Do not operate the unit in flammable, volatile or explosive atmosphere.
6. Do not insert objects through the unit’s intake or outlet grilles. Damage to the ionizer
and/or personal injury may result.
SECTION 5 Installation
5.1 Unpacking
Carefully remove the equipment from the carton and inspect contents. Note that the unit
has been bagged and vacuum sealed in a Class 100 cleanroom. To prevent
contamination and eliminate the need for additional wipe downs, the unit should be un-
bagged in a cleanroom environment.
5.2 Installing
The Centurion single fan ionizer is designed for portable or permanent operation.
Typically, the unit should be positioned such that there is good airflow provided to the
critical area. Generally this requires that the unit be within 36” inches of the intended
target. The stand provided with the unit can be used in a permanent operation by bolting
it to a sturdy flat surface such as a wall or shelf. The unit can be mounted horizontally if
required. The stand may be removed if required.
4