Guidance and manufacturer's declaration - electromagnetic emissions
The Model MKC-300HD is intended for use in the electromagnetic environment specified below. The
customer or the user of the Model MKC-300HD should assure that it is used in such an environment.
Emissions test
Compliance Electromagnetic
The Model MKC-300HD uses RF energy only for its
internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11 Group 1
RF emissions
CISPR 11 Class A
Harmonic emissions
IEC61000-3-2
Class A
Voltage
fluctuations/flicker
emissions IEC61000-3-3
Complies
The Model MKC-300HD is suitable for use in all
establishments other than domestic and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Immunity test
IEC 60601 test level
compliance level
Electromagnetic environment
guidance
Electrostatic
discharge(ESD)
IEC61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC61000-4-4
±2kV for power
supply lines
±1kV for input/output
lines
±2kV for power
supply lines
±1kV for input/output
lines
Main power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC61000-4-5
±1kV differential
mode
±2kV common mode
±1kV differential
mode
±2kV common mode
Main power quality should be
that of a typical commercial or
hospital environment
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines.
IEC61000-4-11
<5% Ut
(>95% dip In Ut)
for 0.5 cycle
40% Ut
(60% dip In Ut)
for 5 cycle
70% Ut
(30% dip In Ut)
for 25 cycle
<5% Ut
(>95% dip In Ut)
for 5 cycle
<5% Ut
(>95% dip In Ut)
for 0.5 cycle
40% Ut
(60% dip In Ut)
for 5 cycle
70% Ut
(30% dip In Ut)
for 25 cycle
<5% Ut
(>95% dip In Ut)
for 5 cycle
Main power quality should be
that of a typical commercial or
hospital environment. If the user
of the MODEL MKC-300HD
requires continued operation
during power mains
interruptions, It is recommended
that the MODEL MKC-300HD be
powered from an uninterruptible
power supply or battery.
Power frequency
(50/60 Hz)
magnetic field
IEC61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at characteristic of a
typical location In a typical
commercial or hospital
environment.
NOTE Ut is the a.c.mains voltage prior to application of the test level.
DN-27466