22

Setting menu parameters

The device can be operated in mode A, B and P1 to P9. The mode 
setting will be stored automatically when operation is over, or the 
device is switched off. User may change the mode setting in the 
main menu if necessary.

• 

Operation mode A

In this mode, when current operation is over or the device is 
switched off, all working settings will not be stored.

• 

Operation mode B

In this mode, when current operation is over or the device is 
switched off, all menu and working settings will be stored.

• 

Operation mode P1 - P9

In mode P1-P9, the user can set and save required parameters,
(target speed, timer value and rev value) in program. The param-
eter settings can only be changed in the menu program. These 
settings can also be used directly when necessary. 

• 

Setting operation mode between A and B

)

  Press the “Menu/Back” button (C) in the main menu. 

)

 Turn the “Rotating/Pressing” knob (B) to select the “mode” item 

in the main menu.

)

  Press the “Rotating/Pressing” knob (B).

)

 Turn the “Rotating/Pressing” knob (B) to select mode A or B in 

submenu.

)

 Press the “Rotating/Pressing” knob (B) again to complete the 

process.

)

  Press the “Menu/Back” button (C) to return to the working screen.

• 

Setting operation mode between P1 and P9

)

 Press the “Menu/Back” button (C) in the main menu. 

)

 Turn the “Rotating/Pressing” knob (B) to select the “mode” 

item in the main menu.

11

Operation mode

11.1

)

 Press the “Rotating/Pressing” knob (B).

)

 Turn the “Rotating/Pressing” knob (B) to select the “mode” 

item in mode P1 and P9 in submenu.

)

 Press the “Rotating/Pressing” knob (B) and enter desired 

menu program.

)

 Turn the “Rotating/Pressing” knob (B) to select desired pro-

gram item.

)

 Press the “Rotating/Pressing” knob (B) .

)

 Turn the “Rotating/Pressing” knob (B) to change the value.

)

 Press the “Rotating/Pressing” knob (B) again to complete the 

setting.

)

 Press the “Menu/back” button (C) to complete the process, 

display will return to working screen.

Note:

 The target speed and timer value in the program should be 

less than “rpm max” and “timer max” setting in menu. Otherwise, 
the device will run according to the setting in the menu. 

Operation language

11.2

The language  can be set to: English, German, Chinese, French, 
Spanish, Italian and Korean. The language setting will be stored au-
tomatically when operation is over or the device is switched off. 
User can change the language setting in the main menu if necessary.

• 

Setting operation language

)

 Press the “Menu/Back” button (C) in the main menu. 

)

 Turn the “Rotating/Pressing” knob (B) to select the “lan-

guage” item in the main menu.

)

  Press the “Rotating/Pressing” knob (B) to confirm “lan-

guage”.

)

 Turn the “Rotating/Pressing” knob (B) to select a language.

)

 Press the “Rotating/Pressing” knob (B) again to complete 

the setting.

)

 Press the “Menu/Back” button (C) to return to the working 

screen.

Содержание ULTRA-TURRAX Tube Drive P control

Страница 1: ...IKA ULTRA TURRAX Tube Drive P control 20000005139 UTTD P control_042018...

Страница 2: ...ommissioning 20 9 Switching on 20 10 Factory setting 21 10 1 Resetting parameters to factory setting RESET 21 11 Set menu parameters 22 11 1 Operation mode 22 11 2 Operation language 22 11 3 Maximum r...

Страница 3: ...the drive unit prior to operating the device The tube must only be attached and removed while the motor is stationary The IKA tubes must always be closed when the device is in operation Switch off the...

Страница 4: ...T 50 Dispersing Stirring 5 15 ml 15 50 ml A B P1 P9 Turbo Rev with Tube BMT 20 with Tube BMT 50 Crushing Mixing wet dry 2 15 ml 15 50 ml A B P1 P9 Turbo Rev with Tube DIS 50 with Tube DIS 300 X Crushi...

Страница 5: ...as described in RS 232 interface Command syntax and format The following applies to the command set Commands are generally sent from the computer Master to the device Slave The device sends only at t...

Страница 6: ...on is pressed after this command is sent to device Remote symbol is displayed RESET Switch to normal operation Device Keypad available again and clear the Re mote symbol IN_PV_X X 4 Read actual speed...

Страница 7: ...ings proceed as following Pull out main plug from the device Hold the controller buttons Menu Back and Timer simultane ously and re insert main plug in power outlet The unit is now reset to the factor...

Страница 8: ...the Rotating Pressing knob B to select the mode item in the main menu 11 Operation mode 11 1 Press the Rotating Pressing knob B Turn the Rotating Pressing knob B to select the mode item in mode P1 an...

Страница 9: ...nge from 10 seconds to 30 minutes This is achieved in magnitudes of 10 seconds In remote control mode via computer adjusting magnitude is 1 second The default setting max rpm is 6000 rpm in the device...

Страница 10: ...v value with the Rotating Pressing knob B the value can be set from 10 to 60 seconds where adjustment magni tude is 1 second Note If both Rev function and Timer function are activated the Rev value sh...

Страница 11: ...designation of the spare part see www ika com spare parts diagram and spare parts list Turbo function will be stopped and rotation speed will return to original setting A symbol appears to indicate th...

Страница 12: ...lts resulting from improper use or insufficient care and mainte nance contrary to the instructions in this operating manual See more accessories on www ika com Repair Please send in device for repair...

Страница 13: ...A Mode B Mode P1 P9 rpm 400 6000 in nitely adjustable adjustment magnitude 10 rpm up to 8000 up to 6000 up to 6000 up to 6000 Speed setting rotating pressing knob Speed display digital Timer 10 secon...

Отзывы: