25

Error codes

13

Maintenance

14

The drive unit is maintenance-free but not immune from wear. The 
carbon brushes of the motor wear down over the time.

For cleaning disconnect the main plug.

Only use cleansing agents which have been recommended by 

IKA

Dirt

 

 

Cleaning agent

Dyes 

 

isopropyl alcohol

Construction materials  water containing tenside/isopropyl alcohol
Cosmetics 

 

water containing tenside/isopropyl alcohol

Foodstuffs 

 

water containing tenside

Fuels 

 

water containing tenside

For materials which are not listed, please request information from 

IKA 

application support.

Wear protective gloves during cleaning the instruments.
Electrical instruments may not be placed in the cleansing agent for 
the purpose of cleaning.
Do not allow moisture to get into the instrument when cleaning.
Before using another than the recommended method for cleaning 
or decontamination, the user must ascertain with 

IKA

 that this 

method does not destroy the instrument.

Spare parts order:

When ordering spare parts, please give:
- machine type
- manufacturing number, see type plate
- item and designation of the spare part, see 

www.ika.com

spare parts diagram and spare parts list.

Ö

  Turbo function will be stopped and rotation speed will return 

to original setting. A symbol   appears to indicate that the 
Turbo function is stopped. Before this symbol disappears, 
turbo function can not be activated again.

Any malfunction during operation may be identified by an error 
message on the display.
Proceed as follows in such cases:

)

 Switch off device using the power switch

)

 Carry out corrective measures

)

 Restart the device.

Error 
code

Cause

Effect

Solution

Error 3

Inside temperature is 
too high

motor off - Switch off the device and 

let it cool down.

- Restart the device.

Error 4

Motor can’t run normal 
(can’t start or not reach 
the target speed)

motor off - Switch off the device.

- Check the structure for 

possible blockades

Error 47
Error 48

Overload

motor off - Switch off the device

- Decrease the speed setting 

or the load

Error 50

Tube is protected in 
order to avoid overload

motor off - Check the tube condition

- Decrease the speed setting 

or the load

If the actions described fails to resolve the fault or another error 
code is displayed then take one of the following steps: 
- Contact the service department 
- Send the device for repair, including a short description of the fault.

 

Содержание ULTRA-TURRAX Tube Drive P control

Страница 1: ...IKA ULTRA TURRAX Tube Drive P control 20000005139 UTTD P control_042018...

Страница 2: ...ommissioning 20 9 Switching on 20 10 Factory setting 21 10 1 Resetting parameters to factory setting RESET 21 11 Set menu parameters 22 11 1 Operation mode 22 11 2 Operation language 22 11 3 Maximum r...

Страница 3: ...the drive unit prior to operating the device The tube must only be attached and removed while the motor is stationary The IKA tubes must always be closed when the device is in operation Switch off the...

Страница 4: ...T 50 Dispersing Stirring 5 15 ml 15 50 ml A B P1 P9 Turbo Rev with Tube BMT 20 with Tube BMT 50 Crushing Mixing wet dry 2 15 ml 15 50 ml A B P1 P9 Turbo Rev with Tube DIS 50 with Tube DIS 300 X Crushi...

Страница 5: ...as described in RS 232 interface Command syntax and format The following applies to the command set Commands are generally sent from the computer Master to the device Slave The device sends only at t...

Страница 6: ...on is pressed after this command is sent to device Remote symbol is displayed RESET Switch to normal operation Device Keypad available again and clear the Re mote symbol IN_PV_X X 4 Read actual speed...

Страница 7: ...ings proceed as following Pull out main plug from the device Hold the controller buttons Menu Back and Timer simultane ously and re insert main plug in power outlet The unit is now reset to the factor...

Страница 8: ...the Rotating Pressing knob B to select the mode item in the main menu 11 Operation mode 11 1 Press the Rotating Pressing knob B Turn the Rotating Pressing knob B to select the mode item in mode P1 an...

Страница 9: ...nge from 10 seconds to 30 minutes This is achieved in magnitudes of 10 seconds In remote control mode via computer adjusting magnitude is 1 second The default setting max rpm is 6000 rpm in the device...

Страница 10: ...v value with the Rotating Pressing knob B the value can be set from 10 to 60 seconds where adjustment magni tude is 1 second Note If both Rev function and Timer function are activated the Rev value sh...

Страница 11: ...designation of the spare part see www ika com spare parts diagram and spare parts list Turbo function will be stopped and rotation speed will return to original setting A symbol appears to indicate th...

Страница 12: ...lts resulting from improper use or insufficient care and mainte nance contrary to the instructions in this operating manual See more accessories on www ika com Repair Please send in device for repair...

Страница 13: ...A Mode B Mode P1 P9 rpm 400 6000 in nitely adjustable adjustment magnitude 10 rpm up to 8000 up to 6000 up to 6000 up to 6000 Speed setting rotating pressing knob Speed display digital Timer 10 secon...

Отзывы: