23
Maximum rotation speed (max rpm)
11.3
Timer maximum value (timer max)
11.4
The default “timer max” is 30 minutes in the device. User can
change the setting in 10 second intervals and up to 30 minutes
in the main menu.
The “timer max” setting will be stored automatically when opera-
tion is over or the device is switched off.
•
Set “timer max”
Press the “Menu/Back” button (C) in the main menu.
Turn the “Rotating/Pressing” knob (B) to select the “timer max”
in the main menu.
Press the “Rotating/Pressing” knob (B)
A mark
will appear to indicate the activated item.
Turn the “Rotating/Pressing” knob (B) to change the value.
Press the “Rotating/Pressing” knob (B) again to complete the
process
Press the “Menu/Back” button (C) to return to the working
screen.
Note:
The timer „timer max” can be adjusted with the “Rotating/
Pressing” knob (B) in a range from 10 seconds to 30 minutes. This
is achieved in magnitudes of 10 seconds. In remote control mode
via computer, adjusting magnitude is 1 second.
The default setting “max rpm” is 6000 rpm in the device. User
can change the setting to the range of 400 to 6000 rpm in the
main menu.
The “max rpm” setting will be stored automatically when opera-
tion is over or the device is switched off.
•
Setting “max rpm”
Press the “Menu/Back” button (C) in the main menu.
Turn the “Rotating/Pressing” knob (B) to select the “max
rpm item in the main menu.
Press the “Rotating/Pressing” knob (B).
A mark
will appear to indicate the activated item.
Turn the “Rotating/Pressing” knob (B) to adjust the value.
Press the “Rotating/Pressing” knob (B) again to complete
the process.
Press the “Menu/Back” button (C) to return to the working
screen.
Note:
The speed „max rpm” can be adjusted with the “Rotating/
Pressing” knob (B) in a range between 400 to 6000 rpm. This is
achieved in magnitudes of 10 rpm. In remote control mode via
computer, adjusting magnitude is 1 rpm.
Содержание ULTRA -TURRAX Tube Drive P control
Страница 41: ...41 BMT 20 BMT 50 D IKA 40 C IKA 3...
Страница 45: ...45 Power A Fig 1 9 A 6000 Timer 30 400 Timer Rev 10 Turbo 8000 Rev Timer Timer...
Страница 50: ...50 16 IKA 24 15 www ika com IKA www ika com...
Страница 65: ...65 IVD BMT 20 BMT 50 IKA 3 40 C IKA...
Страница 73: ...73 13 14 IKA IKA IKA IKA www ika com Error 3 Error 4 Error 47 Error 48 Error 50...
Страница 74: ...74 15 16 www ika com IKA www ika com IKA...
Страница 91: ...91 IVD BMT 20 BMT 50 IKA 40 C IKA BG...
Страница 93: ...93 IVD BMT 20 BMT 50 IKA 40 C IKA EL...
Страница 94: ...94 Note...
Страница 95: ......