26
In accordance with
IKA
®
warranty conditions,
the warranty period is 24 months. For claims un-
der the warranty please contact your local dea-
ler. You may also send the machine direct to our
works, enclosing the delivery invoice and giving
reasons for the claim. You will be liable for freight
costs.
Warranty
Technical data
The warranty does not cover wearing parts, nor
does it apply to faults resulting from improper
use or insuffi cient care and maintenance contra-
ry to the instructions in this operating manual.
Motor data
Rated voltage [VAC]
3 x 400 Y
3 x 230 ∆
3 x 400 Y
3 x 230 ∆
Rated frequency [Hz]
50
50
60
60
Motor speed [rpm]
2880
2850
3520
3520
Power consumption [W]
1943
1943
2381
2381
Power output [W]
1500
1500
1800
1800
Additional data
Speed
rpm 7200
Peripheral speed
m/s
21.9
(generators serie S 65....)
Perm. On- time
%
100
Protection acc. DIN EN 60529
IP 54
Overvoltage category
II
Protection class
I
Contamination level
2
Protection on overload
Overcurrent and undervoltage switch on
the AC motor
Perm. ambient temperature
°C
+ 5 bis + 40
Perm. humidity
%
80
Operating position
On the stand, dispersion tool vertically
downwards
Drive
Vane-cooled AC motor with
Poly-V-belt drive transmission stage
Stand mounting
Flange in lower section of housing
Housing material
Aluminium
Noise level (without disp. tool)
dbA
75
Dimensions (WxDxH)
mm
190x580x380
Weight
kg 28
Operation at a terrestrial altitude
m
max. 2000
Subject to technical changes!
Содержание T 65 D
Страница 2: ......
Страница 8: ...8...
Страница 20: ...20...
Страница 32: ...32...
Страница 52: ...52 30 mm IKA IKA...
Страница 66: ...66 IKA 5 B 30 mm...
Страница 69: ...69 Notes Notes Notes Notes Notes Notes Notes Verdrahtungsplan Wiring diagram Plan de c blage...
Страница 70: ...Notes...
Страница 71: ......