36
Technical data
Error Codes
Any malfunctions during operation will be identified by an error message on the display.
Once a serious error message has been displayed, the lift moves to the top end position and the device can no longer be operated.
Proceed as follows in such cases:
- switch off device using the main switch,
- carry out corrective measures,
- restart device.
Error code
Cause
Effect
Correction
E03
Temperature inside the device
is too high
Room temperature > 40 °C
- Switch off the device
- Let the device cool down
- Check the location
(see Technical data)
E04
Target speed is not reached
Motor blocked
Load too large
Cable break
- Reduce the volume in the evaporator
E09
Flash memory storage error
Transmission error
- Turn the unit off and on again
If the actions described fail to resolve the fault or another error code is displayed, then take one of the following steps:
- contact the service department,
- send the device for repair, including a short description of the fault.
Operating voltage range
Vac
(100 – 240) ± 10%
Rated voltage
Vac
100 – 240
Frequency
Hz
50 / 60
Rated power of RV 8 device(without heating bath)
W
75
Rated power (max.) of switching adaptor
W
90
Speed
rpm
5 – 300
Speed tolerance
Set speed: < 100 rpm
Set speed: ≥ 100 rpm
rpm
%
± 1
± 1
Speed display
mm
digital
Dimensions of visible display area (W x H)
37 x 18
Display
7 segments display
Smooth start
yes
Lift
manual
Stroke
mm
120
Head angle adjustable
0° – 45°
Cooling surface
cm
2
1500
Cooling water flow rate min.
l/h
30
Cooling water flow rate max.
l/h
100
Cooling water pressure max.
bar
1
Perm. On-time
%
100
Perm. ambient temperature
°C
5 – 40
Perm. relative humidity
%
80
Protection acc. to DIN EN 60529
IP 20
Protection class
I
Contamination level
2
Weight (no glassware; no heating bath)
kg
12.5
Dimensions (W x D x H)
mm
595 x 390 x 615
Operation at a terrestrial altitude
m
max. 2000
Subject to technical changes!