15
Set the first point temperature (step 1, e�g� 10 °C) with
“
Pump (+)
” button (
I
) or “
Pump (-)
” button (
J
)� Confirm
the setting by pressing “
OK/Pump
“ button (
H
)�
Left area of the display appears the set value (10 °C) and
the right area of the display appears the actual temperature
value measured� The “
Set
” LED (
B
) flashes briefly�
The unit now starts and controls to the set value� When
the temperature has reached the set value and has become
constant, the “
Set
” LED (
B
) no longer flashes and the fol-
lowing display appears� The temperature measured by the
unit appears on the left hand side�
Input the calibration value from the reference measuring
instrument (e�g� 10�1 °C) with “
Pump (+)
” button (
I
) or
“
Pump (-)
” button (
J
)�
Confirm the value by pressing the “
OK/Pump
“ button (
H
)�
Back to previous screen for input a new value by pressing
“
Temp
” button (
K
)�
Back
OK
The first point calibration is finished now�
Calibration of the other points is performed in the same way�
Example: 2-point calibration:
Dip the temperature sensor of the reference measuring in-
strument into the bath fluid�
Select 2-point calibration in the menu� Press the “
OK/
Pump
“ button (
H
) to start the 2-point calibration�
Reset to factory settings (RST):
RST (No):
Without resetting to factory settings�
RST (Yes):
Resetting to factory settings�