background image

6

7

• 

Unpacking

- Please unpack the device carefully

- In the case of any damage a detailed report must be set im-

mediately (post, rail or forwarder

Unpacking

• Delivery scope

Telescopic stand 

- Telescopic stand corresponding to 
  the type ordered T 653 oder R 474
- Operating instructions

For technical reasons associated with shipping, the telescopic 
stand is supplied stripped down, as pre-assembled modules..

The stand pillar is inserted into the baseplate so that the guide 
groove is at the back. The black handwheel bolt is then on the 
right. The stand pillar is then secured from below with a hexagon 
nut, toothed lock washer and a large shaped washer.

All four rubber feet are height-adjustable, so the stand can be set 
up so it does not rock, even on uneven floors.

In addition, a hole is provided in the baseplate of the R 474 tele-
scopic stand, for screwing on a 16 mm diameter support rod. A 
clamp can be fixed on this rod.

Set-up and commissioning

Mounting the unit

The

 

IKA 

RW 47 D agitator unit or the T 65 D ULTRA-TURRAX

®

 dis-

persion unit are placed on the arm of the stand and bolted on from 
below with the bolts supplied. The help of an assistant is needed 
for screwing on the T 65 D ULTRA-TURRAX

®

 dispersion unit.

The cable clips included in the delivery package are required for 
fixing the power cable of the unit.

Safety disconnection device

The height-adjustable telescopic stand must be fitted with a safe-
ty disconnection device. We recommend installation of the SI 400  
safety limit switch together with the SI 474 clamp support. You will 
find information on their installation in the operating instructions 
of the units.

Commissioning

The stand can be raised and lowered by loosening the handwheel 
bolt. The integral gas shock absorber helps to keep the manual 
force needed for this as light as possible. The manual force must 
be applied close to the stand pillar, so that the stand can be raised 
and lowered with only a slight use of force. If the force for raising 
or lowering is applied at a distance from the stand pillar, a turning 
moment results, leading to increased friction on the pillar guide. As 
a result, the stand can only be moved with great difficulty, or can-
not be moved at all. Besides this, the danger of the stand toppling 
over is increased.

When the unit is at the correct height, the handwheel bolt is tigh-
tened by hand. Check the tightness of the handwheel bolt regular-
ly when using the unit, and retighten it if necessary.

Correct use

The safety of the user cannot be guaranteed if the appliance is 
operated with accessories that are not supplied or recommended 
by the manufacturer or if the appliance is operated improperly 
contrary to the manufacturer’s specifications or if the appliance is 
modified by third parties.

Use

The T 653 telescopic stand is used solely for mounting 

T 65 D 

ULTRA-TURRAX

®

 dispersion units.

The R 474 telescopic stand is used solely for mounting

 IKA 

RW 47 

D agitator units.

Setting up the stand

Source language: German

EN

Page

EU Declaration of conformity 

2

Explication of warning symbols 

6

Safety instructions 

6

Unpacking 7
Correct use 

7

Set-up and Commissioning 

7

Maintenance and cleaning 

8

Accessories 8
Technical data 

8

Warranty 8

Contents

Explication of warning symbols

General hazard

This symbol identifies information 

that is of vital importance for safeguarding your health and safety

Disregarding this information can lead to health impairment and injuries.

This symbol identifies information and notes that are of importance for correct handling of the unit.

Safety instructions

    

• 

Read the operating instructions in full before 

    

  starting up and follow the safety instructions.

     

• Keep the operating instructions in a place where they 

 

  can be accessed by everyone.

• Ensure that only trained staff work with the appliance.
• Follow the safety instructions, guidelines, occupational health 

and safety and accident prevention regulations.

• The stand pillar must be inserted into the baseplate so that the 

guide groove is at the back. Other alignments are not permissib-
le, as otherwise the stand can topple over when a unit is moun-
ted on it.

• 

Set up the stand in a spacious area on an even, stable, clean, non-

slip, dry and fireproof surface.

     

• Any small unevennesses that cause the stand to rock 

 

   must be compensated for with the height-adjustable  

 

   feet.

• The T 65 D ULTRA-TURRAX

®

 dispersion unit must be lifted on to 

the guide head of the telescopic stand with the aid of an assis-
tant, and secured with the two screws and toothed lock washers 
(supplied), in accordance with the T 65 D ULTRA-TURRAX

®

 ope-

rating instructions. The unit must be mounted in front, vertically 
above the baseplate; a different alignment is not permissible, as 
otherwise the assembly could topple over. 

• The overhead stirrer 

IKA 

RW 47 D must be lifted on to the guide 

head of the telescopic stand with the aid of an assistant, and 
secured with the four screws (supplied), in accordance with the 

IKA 

RW 47 D operating instructions. The unit must be mounted 

in front, vertically above the baseplate; a different alignment is 
not permissible, as otherwise the assembly could topple over.

• This telescopic stand must be fitted with a safety disconnection 

device, which stops the unit if the stand is moved upwards when 
the unit is operating.

• Switch the unit off before moving the telescopic stand upwards.
• When lowering the telescopic stand, take care that no parts of 

the body get between the safety switches (danger of crushing).

• Only switch the unit on when the tool (agitator or dispersion 

unit) is in the vessel.

• The handwheel bolt on the stand can come loose due to vibra-

tion, which would lead to the unit moving down. So, check the 
tightness of the handwheel bolt regularly when using the unit, 
and retighten it if necessary.

• Raise or lower the stand (after slackening the handwheel bolt) 

by applying the manual force as close as possible to the stand 
pillar. The further the force is applied from the pillar, the more 
force you need to apply, plus there is an increased risk of the 
equipment toppling over. 

• Secure the vessel against spinning round, by using a clamp, for 

example.

• Beachten Sie die Betriebsanleitung des Dispergiergerätes T 65 D 

ULTRA-TURRAX

®

 

oder des Rührgerätes 

IKA

 RW 47 D!

• 

Check the appliance and accessories beforehand for damage 
each time you use them. Do not use damaged components.

• 

Protect the appliance and accessories from bumps and impacts.

Содержание R 474

Страница 1: ...Betriebsanleitung DE 3 Operating instructions EN 6 Mode d emploi FR 9 Teleskopstativ Telescopic stand Statif t lescopique R 474 T 653...

Страница 2: ...dere Ausrichtung ist nicht zul ssig da sonst der Aufbau kippen k nnte F r diese h henverstellbaren Teleskopstative muss eine Sicher heitsabschaltung angebracht werden die das Ger t stoppt wenn das Sta...

Страница 3: ...Die schwarze Griffschrau be befindet sich dann auf der rechten Seite Mit Sechskantmutter Zahnscheibe und einer gro en gepr gten Scheibe wird die Stativ s ule von unten festgeschraubt Alle vier Gummif...

Страница 4: ...rrect use 7 Set up and Commissioning 7 Maintenance and cleaning 8 Accessories 8 Technical data 8 Warranty 8 Contents Explication of warning symbols General hazard This symbol identifies information th...

Страница 5: ...de dispersion se trouve dans le r cipient Les vibrations peuvent conduire au desserrage de la vis de poig n e sur le statif ce qui peut entra ner l affaissement de l appareil Il faut donc contr ler r...

Страница 6: ...ommand T 653 ou R 474 Mode d emploi Pour des raisons d exp dition le statif t lescopique est d sassem bl par ensembles principaux pr mont s La colonne de statif doit tre ins r e dans le socle de fa on...

Страница 7: ...900 eMail info ika in UNITED KINGDOM IKA England LTD Phone 44 1865 986 162 eMail sales england ika com MALAYSIA IKA Works Asia Sdn Bhd Phone 60 3 6099 5666 eMail sales lab ika my CHINA IKA Works Guan...

Отзывы: