design voltage:
VAC
220 - 230 ±
10%
oder
VAC
115 ±
10%
oder
VAC
100 ±
10%
design frequency:
Hz
50/60
input power total
W
630 (by 230V)
Drive:
EC-motor with ball bearings,
without brushes
Max. power consumption
W
12 / 5
/ output:
speed range:
rpm
50...1500
infinitely variable setting
speed variety (without load,
%
± 10
1500rpm):
stirring quantity (H
2
O):
l
20
Heating plate:
diameter of heating plate:
mm
135
heating power: (230V / 115V)
W
600 ±
10%
(100V)
W
455 ±
10%
temperature range:
°C
RT .... 340 ±20K
Infinitely variable safety
°C
50 - 380 ±20K
temperature “SAFE TEMP”
temperature fluctuation of
K
±1
the surface temperature
control accuracy of medium
K
±1
in ETS-D operation
contact thermometer-
connection
DIN 12 878 class 2
permitted ambient
temperature:
°C
+5 to +40
permitted relative humidity:
%
80
permitted duration of operation:
%
100
prozection class acc. EN 60529:
IP 42
fuses:
F1/F2 T6,3A /250V
Id-Nr 14 080 00
F3 T0,8T /250V
Id-Nr 27 562 00
contamination level:
2
overvoltage category:
II
protection class:
1(protective earth)
operation at a terrestrial altude
m
max 2000 above sea level
dimensions:
mm
160 x 280 x 85
weight:
kg
2,4
14
RET BC 1001
Technical data
RET basic C 1204 14.01.2005 11:44 Uhr Seite 14