Commands:

NAMUR 

Commands

Function

IN_PV_1

Read the external actual temperature

IN_PV_2

Read the internal actual temperature

IN_PV_3

Read the safety actual temperature

IN_PV_4

Read the pump actual speed

IN_SP_1

Read the internal setting temperature (if ET=0)
Read the external setting temperature (if ET=1)

IN_SP_3

Read the safety setting temperature

IN_SP_4

Read the pump setting speed

IN_TMODE

Read temperature control
0: internal regulation
1: external regulation

OUT_SP_1 xxx

Set the internal setting temperature XXX (if ET=0)
Set the external setting temperature XXX (if ET=1)

OUT_SP_12@n

Set the WD safety temperature with echo of the set (defined) value�

OUT_SP_4 xxx

Set the pump speed XXX

OUT_SP_42@n

Set the WD-safety speed with echo of the set (defined) value�

OUT_TMODE_0

Set to Internal temperature control

OUT_TMODE_1

Set to External temperature control

OUT_WD1@n

Start the watchdog mode 1 and set the time for the watchdog to n (20���1500) seconds�
Echo of the Watchdog time�
During a WD1-event, the heating and pump functions are switched off�
This command needs to be send within the watchdog time� 

OUT_WD2@n

Start the watchdog mode 2 and set the watchdog time to n (20���1500) seconds�
Echo of the watchdog time� 
During a WD2-event, the set temperature is changed to the WD safety temperature and the pump set 
speed is set to the WD safety speed�
This command needs to be send within the watchdog time� 

RESET

Reset the PC control and stop the device functions�

START_1

Start the tempering function

START_4

Start the pump function

STOP_1

Stop the tempering function

STOP_4

Stop the pump function

PC 1�1 Cable:

 This cable is required to connect RS 232 interface (

11

) to a PC�

1
2 RxD

3 TxD

4

5 GND

6

7 RTS

8 CTS

9

1

RxD 2

TxD 3

4

GND 5

6

RTS 7

CTS 8

9

PC

1  

2  

3  

4  

5

6  

8  

9

9  

8  

7  

6

5  

4  

3  

2  

1

Fig. 14

Connections between device and external devices:

Содержание CBC 5 basic

Страница 1: ...20000013102 IKA CBC 5 basic CBC 5 basic_112016 Operating instructions EN...

Страница 2: ...er 4 Handle 5 Venting grid 6 Mains switch 7 Power socket 8 Cable clip 9 Pump connection IN 10 Pump connection OUT 11 RS 232 interface 12 USB interface 13 External temperature sensor socket Device setu...

Страница 3: ...panel and display 10 Setting the safety temperature 11 Commissioning 11 Menu settings 12 Menu structure 12 Basic guide to setting the menu options 13 Operating mode MODE 13 Maximum temperature HI T 13...

Страница 4: ...10 2 051 EN 61326 1 EN 60529 and DIN 12876 1 Indicates an extremly hazardous situation which if not avoided will result in death serious injury Indicates a hazardous situation which if not avoided can...

Страница 5: ...For example users can implement measures that inhibit fire or explo sions or comprehensive monitoring equipment Process pathogenic material only in closed vessels under a suitable fume hood Please co...

Страница 6: ...at also supplies residential areas The safety of the user cannot be guaranteed if the device is operated with accessories that are not sup plied or recommended by the manufacturer if the device is ope...

Страница 7: ...ources and do not expose to sun light Cooling machine pump motor and electronics produce intrinsic heat that is dissipated via the venting grids 5 Never cover these venting grids Note After setting up...

Страница 8: ...valve is closed Ro tate clockwise to the stop position see Fig 7 Note Please note information in chapter Commissioning Connect the mains plug and turn on the device with mains switch 6 The low level...

Страница 9: ...distilled or bi distilled water are unsuitable for operation due to corrosive properties of these media High purity water and distillates are suitable as a medium after adding 0 1 g soda Na2 CO3 sodi...

Страница 10: ...speed value Start stop the pump function K Rotating pressing knob Set the temperature value Start stop the heating refrigerating function Navigation selecting the settings in the menu Set and confirm...

Страница 11: ...and fan starts to run at once After pressing the mains switch 6 the start up of the com pressor is delayed about 3min In working status press the right knob J to stop the pump function The heating ref...

Страница 12: ...play will change back and show pump speed When the heating refrigerating function is activated you can enter the menu option by pressing menu key G about 3 seconds Menu settings MODE Maximum temperatu...

Страница 13: ...luid fluid type Note The maximum value is limited by the set safety temperature Minimum temperature LO T The minimum adjustable value 25 C Maximum speed HI R The maximum adjustable value 4600 rpm Mini...

Страница 14: ...not displayed in the menu list The sampling time EuT 2 s is the interval over which the external control deviation and the respective control variable dependent on Kp 2 Ti 2 Td 2 Df t and FAST is dete...

Страница 15: ...cs of the interface and the al location of signal status standard RS 232 applies in accor dance with DIN 66 259 Part 1 Transmission procedure asynchronous character transmis sion in start stop mode Ty...

Страница 16: ...ODE_1 Set to External temperature control OUT_WD1 n Start the watchdog mode 1 and set the time for the watchdog to n 20 1500 seconds Echo of the Watchdog time During a WD1 event the heating and pump f...

Страница 17: ...erial Water containing tenside isopropyl alcohol Cosmetics Water containing tenside isopropyl alcohol isopropyl alcohol Water containing tenside Foodstuff Water containing tenside Fuel Water containin...

Страница 18: ...off Heating refrigerating off Device internal temperature is too high Check the ambient temperature and let the device cool down Err 10 Pump off Heating refrigerating off PC communication failure Chec...

Страница 19: ...be nominal width 12 mm H PUR 8 PUR tube nominal width 8 mm H PUR 12 PUR tube nominal width 12 mm H FKM 8 FKM tube nominal width 8 mm H FKM 12 FKM tube nominal width 12 mm Tubing insulations ISO 8 Insu...

Страница 20: ...play LED Heating capacity W 2500 230 VAC 1250 115 VAC Cooling capacity according to DIN 12876 at 4600 rpm 20 C 10 C 0 C 10 C 20 C W 350 400 W at 3200 rpm 320 370 W at 3200 rpm 270 320 W at 3200 rpm 19...

Страница 21: ...IKA warranty conditions the warranty period is 24 months For claims under the warranty please contact your local dealer You may also send the machine direct to our factory enclosing the delivery invo...

Страница 22: ...IKA Werke GmbH Co KG Janke Kunkel Str 10 D 79219 Staufen Tel 49 7633 831 0 Fax 49 7633 831 98 sales ika de www ika com...

Отзывы: