Pressing the lock key (I) for about 2 seconds, the device con-
trol elements can be locked to prevent accidental changes 
that can be done during operation (LED (H) will light up)�
To unlock the control elements, press the lock key again for 
about 2 seconds (LED (H) goes out)� 

Note: 

When the heating function is activated, you can re-

view the external/internal actual temperature on the right 
of the display (speed display) by pressing the menu key (G)� 
For details, see following table:

ET=0

ET=1

Temp. display

rpm display 
(pressing menu 
key (G))

activated

internal control 
temperature

external actual 
temperature

activated

external control 
temperature

internal actual 
temperature

When 

ET

=0, if external sensor is not connected, when 

pressing menu key (G), the right display shows “- - - -”�

After pressing menu key (G) about 10 seconds, value in the 
right display will change back and show pump speed�

When the heating/refrigerating function is activated, you 
can enter the menu option by pressing menu key (G) about 
3 seconds�

Menu settings

MODE

Maximum temperature (HI T)
Minimum temperature (LO T)
Maximum speed (HI R)
Minimum speed (LO R)
Fluid type (FLUI)

The type of temperature PID control arithmetic (AUTO)
(heating)

External and internal temperature control (ET)
(heating)

Alarm and key tone (BEEP)

Cooling (COOL)

Menu

• 

Menu structure:

Default settings

A
B
C

Proportional coefficient of PID (Kp 1)
Integrate time of PID (Ti 1)
Differential time of PID (Td 1)

AUTO 1
AUTO 0

External sampling time (EuT 2)
Proportional coefficient of PID (Kp 2)
Integrate time of PID (Ti 2)
Differential time of PID (Td 2)
DF t 

FASt 0
FASt 1

ET 0
ET 1

BEEP 0
BEEP 1

0
1

activated
-
-

200 °C
-25 °C
4600 rpm
1000 rpm
8

activated
15�0
5�00
5�0

activated
90 s
1�0
1�5
0�0
100�0
-
activated

-
activated

-
activated

activated
-
-

 Note:

 

DF t: maximum temperature difference between internal and external temperature�

Calibration (CALI)

CALI 0
CALI 2
CALI 3

Содержание CBC 5 basic

Страница 1: ...20000013102 IKA CBC 5 basic CBC 5 basic_112016 Operating instructions EN...

Страница 2: ...er 4 Handle 5 Venting grid 6 Mains switch 7 Power socket 8 Cable clip 9 Pump connection IN 10 Pump connection OUT 11 RS 232 interface 12 USB interface 13 External temperature sensor socket Device setu...

Страница 3: ...panel and display 10 Setting the safety temperature 11 Commissioning 11 Menu settings 12 Menu structure 12 Basic guide to setting the menu options 13 Operating mode MODE 13 Maximum temperature HI T 13...

Страница 4: ...10 2 051 EN 61326 1 EN 60529 and DIN 12876 1 Indicates an extremly hazardous situation which if not avoided will result in death serious injury Indicates a hazardous situation which if not avoided can...

Страница 5: ...For example users can implement measures that inhibit fire or explo sions or comprehensive monitoring equipment Process pathogenic material only in closed vessels under a suitable fume hood Please co...

Страница 6: ...at also supplies residential areas The safety of the user cannot be guaranteed if the device is operated with accessories that are not sup plied or recommended by the manufacturer if the device is ope...

Страница 7: ...ources and do not expose to sun light Cooling machine pump motor and electronics produce intrinsic heat that is dissipated via the venting grids 5 Never cover these venting grids Note After setting up...

Страница 8: ...valve is closed Ro tate clockwise to the stop position see Fig 7 Note Please note information in chapter Commissioning Connect the mains plug and turn on the device with mains switch 6 The low level...

Страница 9: ...distilled or bi distilled water are unsuitable for operation due to corrosive properties of these media High purity water and distillates are suitable as a medium after adding 0 1 g soda Na2 CO3 sodi...

Страница 10: ...speed value Start stop the pump function K Rotating pressing knob Set the temperature value Start stop the heating refrigerating function Navigation selecting the settings in the menu Set and confirm...

Страница 11: ...and fan starts to run at once After pressing the mains switch 6 the start up of the com pressor is delayed about 3min In working status press the right knob J to stop the pump function The heating ref...

Страница 12: ...play will change back and show pump speed When the heating refrigerating function is activated you can enter the menu option by pressing menu key G about 3 seconds Menu settings MODE Maximum temperatu...

Страница 13: ...luid fluid type Note The maximum value is limited by the set safety temperature Minimum temperature LO T The minimum adjustable value 25 C Maximum speed HI R The maximum adjustable value 4600 rpm Mini...

Страница 14: ...not displayed in the menu list The sampling time EuT 2 s is the interval over which the external control deviation and the respective control variable dependent on Kp 2 Ti 2 Td 2 Df t and FAST is dete...

Страница 15: ...cs of the interface and the al location of signal status standard RS 232 applies in accor dance with DIN 66 259 Part 1 Transmission procedure asynchronous character transmis sion in start stop mode Ty...

Страница 16: ...ODE_1 Set to External temperature control OUT_WD1 n Start the watchdog mode 1 and set the time for the watchdog to n 20 1500 seconds Echo of the Watchdog time During a WD1 event the heating and pump f...

Страница 17: ...erial Water containing tenside isopropyl alcohol Cosmetics Water containing tenside isopropyl alcohol isopropyl alcohol Water containing tenside Foodstuff Water containing tenside Fuel Water containin...

Страница 18: ...off Heating refrigerating off Device internal temperature is too high Check the ambient temperature and let the device cool down Err 10 Pump off Heating refrigerating off PC communication failure Chec...

Страница 19: ...be nominal width 12 mm H PUR 8 PUR tube nominal width 8 mm H PUR 12 PUR tube nominal width 12 mm H FKM 8 FKM tube nominal width 8 mm H FKM 12 FKM tube nominal width 12 mm Tubing insulations ISO 8 Insu...

Страница 20: ...play LED Heating capacity W 2500 230 VAC 1250 115 VAC Cooling capacity according to DIN 12876 at 4600 rpm 20 C 10 C 0 C 10 C 20 C W 350 400 W at 3200 rpm 320 370 W at 3200 rpm 270 320 W at 3200 rpm 19...

Страница 21: ...IKA warranty conditions the warranty period is 24 months For claims under the warranty please contact your local dealer You may also send the machine direct to our factory enclosing the delivery invo...

Страница 22: ...IKA Werke GmbH Co KG Janke Kunkel Str 10 D 79219 Staufen Tel 49 7633 831 0 Fax 49 7633 831 98 sales ika de www ika com...

Отзывы: