Technical Data
S 25 KD
(- LR)
- 18 G
S 25 KD
(- LR)
- 25 G
S 25 KD
(- LR)
- 25 F
S 25 KD
(- LR)
- 18 G - S
T
S 25 KD
(- LR)
- 25 G - S
T
Working range
ml
10 - 1500 50 - 2000 100 - 2000 10 - 1500 50 - 2000
Stator/Rotor (Ø)
mm
18/12.6
25/17
25/18
18/13.4
25/20
Gap between rotor and stator
mm
0.3
0.5
0.5
0.25
0.5
Max. allowable speed
rpm
25000
Max. circumferential speed
m/s
16.5
22.2
23.6
17.5
26.2
Min./max. immersion depth
mm
40/185
Material in contact with medium
FFKM and PTFE (conform to the FDA, Food and Drug
Administration rules), AISI 316L
pH range
2 - 13
Suitable for solvents
Yes
Suitable for abrasive substances
No
Max. working temperature
ºC
150
Sterilization methods
all meththods (except the ball bearing)
Working range vacuum
mbar
100
Ultimate fineness, suspensions
μm
10 - 50
15 - 50
5 - 25
10 - 50
15-50
Ultimate fineness, emulsions
μm
1 - 10
1 - 10
1 - 5
1 - 10
1 - 10
Subject to technical changes!
Warranty
In accordance with
IKA
warranty conditions, the warranty period is 24
months. For claims under the warranty please contact your local dealer.
You may also send the machine direct to our works, enclosing the delivery
invoice and giving reasons for the claim. You will be liable for freight costs.
The warranty does not cover wearing parts, nor does it apply to faults
resulting from improper use or insufficient care and maintenance contrary
to the instructions in this operating manual.
• Pull the shaft (
2
) out of the shaft tube (
3
) in a downward direction. The
individual parts threaded on the shaft (
4
,
5
and
6
) may simply be pulled off.
• The sealing (
5
) shall be replaced in the case of wear.
Note:
The sealing (
5
) mustn’t be assembled in the opposite direction (see
Fig. 3
)!
The assembly of the dispersion tool takes place in re-
verse sequence.
The rotor must not be tightened too strongly, otherwise the thread will
be damaged.
Spare parts order:
Spare parts diagram see front page.
For orders of spare parts please specify the serial number on the shaft
tube, the type of dispersion tool and the designation of the spare part.
Repair:
Please send the dispersion tool for repair only after it has been
cleaned and is free from any materials which may constitute a
health hazard.
For repair, please request the “
Decontamination Certificate
” form
IKA
or use the download printout of it from
IKA
website:
www.ika.com
.
If you require servicing, return the instrument in its original packaging. Stor-
age packaging is not sufficient. Please also use suitable transport packaging.
NOTICE