16
Power supply
Battery powered mode:
6VDC 4*AAA alkaline batteries
Display mode
Blue LCD with white backlight V.A.60mm×40.5mm
Measurement mode
Oscillographic testing mode
Measurement range
Rated cuff pressure:
0mmHg~299mmHg(0kPa ~ 39.9kPa)
Measurement pressure: SYS: 60mmHg~230mmHg
(8.0kPa~30.7kPa) DIA: 40mmHg~130mmHg
(5.3kPa~17.3kPa) Pulse value: (40-199)beat/minute
Accuracy
Pressure: 5°С-40°Сwithin±0.4kPa(3mmHg)
Pulse value: ±5%
Normal working condition
A temperature range of :+5°C to +40°C
A relative humidity range of 15% to 90%,
non-condensing, but not requiring a water vapour
partial pressure greater than 50 hPa An atmospher-
ic pressure range of : 700 hPa to 1060 hPa
Storage & transportation
condition
Temperature:-20°C to +60°C
A relative humidity range of ≤ 93%, non-condens
-
ing, at a water vapour pressure up to 50hPa
Measurement perimeter of
the upper arm
About 22cm~32cm or 22cm~42cm
Net Weight
Approx.182g(Excluding the dry cells)
External dimensions
Approx.110mm×110mm×41mm
Attachment
4×AAA alkaline batteries,user manual
Mode of operation
Continuous operation
Degree of protection
Type BF applied part
Protection against ingress
of water
IP21 It means the device could protected against
solid foreign objects of 12.5mm and greater, and
protect against vertcally falling water drops.
Device Classification
Battery Powered Mode:
Internally Powered ME Equipment
WARNING:
No modification of this equipment is allowed.
Conformance to standards: DSTU EN 1060-1:2015 “Non-Invasive Sphygmomanometer.”
General Requirements. Paragraph 9.2. DSTU EN 1060-3:2015 “Non-Invasive
Sphygmomanometer.” Part 3. Paragraph 9.2.
WARRANTY
Your Automatic Blood Pressure Monitor Model iTop is warranted for 5 years from