8. Measurement range:

Cuff pressure: 0-300 mmHg

Systolic: 60-260 mmHg

Diastolic: 40-199 mmHg

Pulse rate: 40-180 beats/minute

9. Accuracy:

Pressure: ±3 mmHg

Pulse rate: ±5%

10. Environmental temperature for operation: 10°C-40°C(50°F -104°F)

11. Environmental humidity for operation:

85%RH

12. Environmental temperature for storage and transport:-20°C-55°C(-4°F-131°F)

13. Environmental humidity for storage and transport:

85%RH

14. Environmental pressure: 80kPa-105kPa

15. Battery life: more than 80 measurements on a full charge

16. The blood pressure measurement system includes accessories: pump, valve, cuff, and sensor.

Note: These specifications are subject to change without notice.

GENERAL SAFETY AND PRECAUTIONS

1. Read all of the information in the User Guide and other provided instructions before operating the unit.

2. Consult your physician for any of the following situations:

a) The application of the cuff over a wound or inflamed area.

b) The application of the cuff on any limb with intravascular access or therapy, or an arteriovenous (A-V) shunt.

c) The application of the cuff on the arm on the side of a mastectomy.

d) Simultaneous use with other medical monitoring equipment on the same limb.

e) The blood circulation of the user needs to be checked.

3. Do not use this product in a moving vehicle as this may result in inaccurate measurements.

4. Blood pressure measurements determined by this product are equivalent to those obtained by professional healthcare

practitioners using the cuff/stethoscope auscultation method within the limits prescribed by the American National Standard,

Electronic or Automated Sphygmomanometer. This device is also clinically validated according to the 2010 Protocol of the

European Society of Hypertension (ESH 2010).

6. If an Irregular Heartbeat (IHB) is detected during the measurement procedure, the IHB symbol will be displayed in the

“iHealth MyVitals” APP. Under this condition, the Wireless Blood Pressure Wrist Monitor can keep functioning, but the results

may be inaccurate. Please consult your physician for accurate assessment.

The IHB symbol will be displayed under 2 conditions:

1) The coefficient of variation (CV) of pulse period >25%.

2) The difference of adjacent pulse period is

0.14s and more than 53 percent of the total number of pulses readings falls within

this definition.

7. Please do not use any cuff other than that supplied by the manufacturer as this may result in inaccurate measurements.

8. For information regarding potential electromagnetic or other interference between the blood pressure monitor and other

devices together with advice regarding avoidance of such interference, please see ELECTROMAGNETIC COMPATIBILITY

INFORMATION. It is suggested that the blood pressure monitor should be operated at least 10 meters away from electric or

wireless devices (e.g. routers, microwave oven, etc.)

9. If the blood pressure measurement (systolic or diastolic) is outside the rated range specified in part SPECIFICATIONS, the

monitor will immediately display a technical alarm on the screen. In this case, repeat the measurement ensuring that the proper

measurement procedures are followed and/or consult with your medical professional. The technical alarm is preset in the factory

Содержание View

Страница 1: ...T UP REQUIREMENTS 2 SET UP PROCEDURES 3 MEASUREMENT PROCEDURES 3 SPECIFICATIONS 4 GENERAL SAFETY AND PRECAUTIONS 5 BATTERY HANDLING AND USAGE 6 TROUBLESHOOTING 6 CARE AND MAINTENANCE 7 WARRANTY INFORM...

Страница 2: ...gned to measure the systolic and diastolic blood pressures and pulse rate of an adult individual by using a technique in which an inflatable cuff is wrapped around the wrist The measurement range of t...

Страница 3: ...sure measurements 1 Sit comfortably with your feet flat on the floor without crossing your legs Stay still during measurement Do not move your wrist body or the monitor 2 Place your hand palm side up...

Страница 4: ...ar and then select it to pair This may take up to 30 seconds and your monitor s Bluetooth indicator will light up when a connection has been established Launch the iHealth MyVitals app to sync with yo...

Страница 5: ...ethod within the limits prescribed by the American National Standard Electronic or Automated Sphygmomanometer This device is also clinically validated according to the 2010 Protocol of the European So...

Страница 6: ...ery when the battery is less than 25 charged Overcharging the battery may reduce its lifetime When in charging mode the charging status will be displayed on the LED screen See the table below for deta...

Страница 7: ...gnetic interference Press the START STOP button and hold for about 10 seconds to reset the device CARE AND MAINTENANCE 1 If this monitor is stored near freezing temperatures allow it to return to room...

Страница 8: ...lue The sign graphical symbol white Symbol for ENVIRONMENT PROTECTION Waste electrical products should not be disposed of with household waste Please recycle where facilities exist Check with your loc...

Страница 9: ...easures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receive...

Страница 10: ...c discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be a...

Страница 11: ...all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellul...

Страница 12: ...rs m can be determined using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufactu...