Table 3

For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING

Guidance and manufacturer’s declaration - electromagnetic immunity

BP7S is intended for use in the electromagnetic environment specified below. The user of BP7S should ensure that it is used in such an

environment.

IMMUNITY test

IEC 60601test level

Complian

ce level

Electromagnetic environment - guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms 150 kHz to

80 MHz

3 V/m 80 MHz to 2.5

GHz

3 V

3 V/m

Portable and mobile RF communications equipment should be used no

closer to any part of BP7S, including cables, than the recommended

separation distance calculated from the equation applicable to the

frequency of the transmitter.

Recommended separation distance

80 MHz to 800 MHz

800 MHz to 2.5 GHz

Where P is the maximum output power rating of the transmitter in watts (W)

according to the transmitter manufacturer and d is the recommended

separation distance in meters (m).

Field strengths from fixed RF transmitters, as determined by an

electromagnetic site survey,

a

should be less than the compliance level in

each frequency range.

b

Interference may occur in the vicinity of equipment

marked with the following symbol:

P

d

2

.

1

=

P

d

2

.

1

=

P

d

3

.

2

=

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from

structures, objects and people.

a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur

radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic

environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the

location in which BP7S is used exceeds the applicable RF compliance level above, BP7S should be observed to verify normal operation. If

abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating BP7S.

b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Table 4

For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING

Recommended separation distances between

portable and mobile RF communications equipment and the Wireless Blood Pressure Wrist Monitor

BP7S is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The user of BP7S can help

prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment

(transmitters) and BP7S as recommended below, according to the maximum output power of the communication equipment.

Rated maximum output

power of transmitter

Separation distance according to frequency of transmitter

m

Содержание View

Страница 1: ...T UP REQUIREMENTS 2 SET UP PROCEDURES 3 MEASUREMENT PROCEDURES 3 SPECIFICATIONS 4 GENERAL SAFETY AND PRECAUTIONS 5 BATTERY HANDLING AND USAGE 6 TROUBLESHOOTING 6 CARE AND MAINTENANCE 7 WARRANTY INFORM...

Страница 2: ...gned to measure the systolic and diastolic blood pressures and pulse rate of an adult individual by using a technique in which an inflatable cuff is wrapped around the wrist The measurement range of t...

Страница 3: ...sure measurements 1 Sit comfortably with your feet flat on the floor without crossing your legs Stay still during measurement Do not move your wrist body or the monitor 2 Place your hand palm side up...

Страница 4: ...ar and then select it to pair This may take up to 30 seconds and your monitor s Bluetooth indicator will light up when a connection has been established Launch the iHealth MyVitals app to sync with yo...

Страница 5: ...ethod within the limits prescribed by the American National Standard Electronic or Automated Sphygmomanometer This device is also clinically validated according to the 2010 Protocol of the European So...

Страница 6: ...ery when the battery is less than 25 charged Overcharging the battery may reduce its lifetime When in charging mode the charging status will be displayed on the LED screen See the table below for deta...

Страница 7: ...gnetic interference Press the START STOP button and hold for about 10 seconds to reset the device CARE AND MAINTENANCE 1 If this monitor is stored near freezing temperatures allow it to return to room...

Страница 8: ...lue The sign graphical symbol white Symbol for ENVIRONMENT PROTECTION Waste electrical products should not be disposed of with household waste Please recycle where facilities exist Check with your loc...

Страница 9: ...easures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receive...

Страница 10: ...c discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be a...

Страница 11: ...all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellul...

Страница 12: ...rs m can be determined using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufactu...

Отзывы: