Please visit the support section of the www.ihealthlabs.eu website to see our comprehensive Frequently 
Asked Questions on blood pressure.

9. GENERAL SAFETY AND PRECAUTIONS

1. Carefully read the user guide before operating the device.

2. Consult your physician for any of the following situations:

  a) The application of the cuff over a wound or inflamed area.

  b) The application of the cuff on the arm on the side of a mastectomy.

  c) Simultaneous use with other medical monitoring equipment on the same limb.

  d) The application of the cuff on any limb with intravascular access or therapy, or an arteriovenous (A-V) 

shunt.                                      

  e) The blood circulation of the user needs to be checked.

3. Do not use this product in a moving vehicle as this may result in inaccurate measurements.

4. The IHB symbol will be displayed under 2 sets of circumstances:

  a) The coefficient of variation (CV) of pulse period >25%.

  b) The difference of adjacent pulse period is ≥0.14s and more than 53% of the total number of pulses 

readings falls within this definition.

5. Please do not use any cuff other than that supplied by the manufacturer as this may result in inaccurate 

measurements.

6. The monitor will show 'HI' or 'Lo' as technical alarm on LCD with no delay if the determined blood 

pressure (systolic or diastolic) is outside the rated range specified in part SPECIFICATIONS. In this case, you 

should consult a physician or check if your operation violated the instructions. The technical alarm 

condition (outside the rated range) is pre-set in the factory and cannot be adjusted or inactivated. This 

alarm condition is assigned as low priority according to IEC 60601-1-8. The technical alarm is non-latching 

and needs no reset. The signal displayed on LCD will disappear automatically after about 8 seconds.

             This Monitor is designed for adults and should never be used on infants, young children or patients 

10

Содержание Track KN550-BT

Страница 1: ...ic iHealth Track Connected Arm Blood Pressure Monitor USER MANUAL Note This is a shortened practical version of the user manual For the full version of this user manual please visit www ihealthlabs eu...

Страница 2: ...e 2 Sit comfortably at a table when measuring your blood pressure 3 The cuff should be placed at the same level as your heart 4 Do not talk or move during a measurement 5 Use the same arm every time 6...

Страница 3: ...ure your blood pressure but when you connect it to the free iHealth MyVitals App you will be able to track store manage and share your results with your family doctor or care givers Simply download iH...

Страница 4: ...ttention to polarity c Close the battery cover When the display shows the battery symbol replace all the batteries with new ones Rechargeable batteries are not suitable for this monitor The monitor th...

Страница 5: ...cro fastener d The cuff should fit comfortably around your arm You should be able to insert one finger between your arm and the cuff Note If the cuff becomes dirty remove it from the monitor and wipe...

Страница 6: ...ished the blood pressure and pulse rate will be displayed on the display The blood pressure will be displayed in different colours green for normal yellow for slightly elevated and red for high accord...

Страница 7: ...tton again to review the next measure ment During display the memories keep on pressing button M for three seconds all results in the current memory bank will be deleted If no result stored LCD will s...

Страница 8: ...Health MyVitals App on your smartphone ensure that your Bluetooth is activated in your settings 3 Press the M Memory button on the blood pressure monitor The number of readings in the memory will be d...

Страница 9: ...ure above 15 mmHg Incorrect operation or strong electromagnetic interference EEPROM accessing error Pneumatic system leakage or cuff is too loose during inflation Fail to detect systolic pressure Cuff...

Страница 10: ...he coefficient of variation CV of pulse period 25 b The difference of adjacent pulse period is 0 14s and more than 53 of the total number of pulses readings falls within this definition 5 Please do no...

Страница 11: ...a long time please remove the batteries 6 It is recommended that the performance of this monitor be checked every 2 years 7 Please clean monitor with a soft dry cloth Do not use any abrasive or volat...

Страница 12: ...the product please visit www ihealthlabs eu and consult our Frequently Asked Questions on blood pressure in our support section If you do not find the answers you are looking for please send us a quer...

Страница 13: ...althlabs eu www ihealthlabs eu ANDON HEALTH CO LTD Add No 3 Jinping Street YaAn Road Nankai District Tianjin 300190 China Tel 86 22 60526161 iHealth is a trademark of iHealth Lab Inc The Bluetooth wor...

Страница 14: ...ZE AAA batteries Measurement range Cuff pressure 0 300mmHg Systolic 60 260mmHg Diastolic 40 199mmHg Pulse rate 40 180 beats minute Measuring accuracy Pressure 3mmHg Pulse rate 5 Environmental temperat...

Отзывы: