10
7 Electrical connection
The unit must be connected by a qualified electrician�
►
Disconnect power before connecting the device�
►
Observe the national and international regulations for the installation of
electrical equipment�
►
Ensure voltage supply to EN 50178, SELV, PELV�
►
Connect the device according to the indicated pin connection�
►
A total current consumption of the device of 3 A must not be exceeded�
NOTE!
The IP rating indicated in the data sheet is only guaranteed if the
M12 connectors are firmly screwed�
The unit can be damaged by insufficiently tightened M12
connectors�
►
Firmly screw the M12 connectors to the device�
Accessories and further information:
www�ifm�com
Accessories
ifm article number
Protective cap
E73004
Torque wrench
E70390
7.1 AUX voltage supply
►
Connect the device to the voltage supply using an M12 connection cable�
3
1
2
4
5
Pin
Connection
1
24 V DC
2
not used
3
0 V
4
not used
5
not used