41
UK
11.4.9 Set response of the outputs in case of a fault
► Select [FOU1] / [FOU2] and set the value:
[On] = Output switches ON in case of a fault.
Analogue output switches to a value > 21 mA in case of a fault.
[OFF] = Switching output switches OFF in case of a fault.
Analogue output switches to a value < 3.6 mA in case of a fault.
Examples of faults: defective hardware, signal quality too low.
Overflow is not considered to be a fault!
11.4.10 Set damping for the measured signal
►
Select [dAP] and set damping in seconds;
setting range: 0.0...60.0 s
More information: (→ 6.3.4)
11.4.11 Set delay time in case of a fault
►
Select [dFo] and set a value between 0...10.0 s.
[dFo] only effective in case of a fault. Mind the dynamics of your application.
In case of fast level changes it is recommended to adapt the value step by
step.
More information: (→ 6.3.6)
11.5 Reset all parameters to factory setting
►
Select [rES].
►
Press [Enter] until [rES] is aligned right.
►
Press and hold [
▲
] or [
▼
] until [----] is displayed.
►
Press [Enter].
> The unit reboots and the factory settings are restored.
Note: On delivery the unit is not operational. First, the first set-up must be
made