Integrate the pipe length so that the direction of flow (1) and the
marking arrow (2) show in the same direction.
In case of turbulences at the inflow side additional
baffled pipes
(B)
are recommended:
Mounting position:
• Permitted mounting positions: pipe length vertical, any position
(fig. 1, 2), pipe length horizontal, unit vertical (fig. 3, 4), unit on
side, pipe length left (fig. 5).
• Avoid the mounting position shown in fig. 6 (unit on side, pipe
length right). If the flow rate is low, the specified measurement
accuracy cannot be adhered to.
28
90° elbow
B = 15 x pipe diameter (D)
two 90° elbows, one plane
B = 20 x pipe diameter (D)
two 90° elbows, two planes
B = 25 x pipe diameter (D)
valve, slide
B = 45 x pipe diameter (D)
1
2
3
5
6
4