- 39 -
3.2.
Accuracy specifications
Accuracy is specified at 23°C ± 5°C, <75% RH.
All accuracy is specified as ±[(% of reading) + (value of least
significant digit)].
For most precise measurements, the meter has to be zeroed first.
Resistance:
Range Resolution
100/120Hz 1kHz 10kHz 100kHz
20.000
Ω
0.001
Ω
─
1.0%+3
1.0%+3
2.0%+3
200.00
Ω
0.01
Ω
1.0%+3
0.2%+2
0.3%+2
0.6%+3
2.0000 k
Ω
0.0001
k
Ω
0.3%+2 0.2%+2
0.3%+2
0.6%+3
20.000 k
Ω
0.001
k
Ω
0.3%+2
0.2%+2
0.3%+2
0.6%+3
200.00 k
Ω
0.01
k
Ω
0.5%+2
0.5%+2
0.5%+2
1.0%+3
2.0000 M
Ω
0.0001
M
Ω
1.0%+3 1.0%+3 1.0%+3
─
2.000M
Ω
0.001
M
Ω
─
─
─
2.0%+3
20.000 M
Ω
0.001
M
Ω
2.0%+3 2.0%+3
─
─
20.00 M
Ω
0.01
M
Ω
─
─
2.0%+3
─
200.0 M
Ω
0.1
M
Ω
2.0%+3
2.0%+3
─
─
Capacitance:
Range Resolution
100/120Hz 1kHz 10kHz 100kHz
200.00 pF
0.01 pF
─
─
1.2%+5
2.0%+5
2000.0 pF
0.1 pF
─
2.0%+3
0.3%+2
0.6%+3
20.000 nF
0.001 nF
2.0%+3
0.2%+2
0.3%+2
0.6%+3
200.00 nF
0.01 nF
0.3%+2
0.2%+2
0.3%+2
0.6%+3
2000.0 nF
0.1 nF
0.3%+2
0.2%+2
0.6%+2
2.0%+5
20.000
μ
F 0.001
μ
F 0.3%+2
0.6%+2
1.2%+5
─
20.00
μ
F 0.01
μ
F
─
─
─
3.0%+5
(10
μ
F max.)
200.00
μ
F 0.01
μ
F 0.6%+2
1.0%+3
─
─
200.0
μ
F 0.1
μ
F
─
─
3.0%+5
(100
μ
F max.)
─
2000.0
μ
F 0.1
μ
F 1.0%+3
─
─
─
2000
μ
F 1
μ
F
─
1.2%+3
─
─
20.00 mF
0.01 mF
1.2%+3
─
─
─
*If reading <2000, unit on display is pF