To access the Relative mode proceed as follows:
1.
Zero the meter. See pages 14-16 for details.
2.
Select the parameter to be tested.
L
,
C
,
R
, or
DCR
3.
Attach the chosen standard to the test points.
4.
Press the
REL%
key.
The meter should now display the
Δ
symbol.
5.
Remove the standard and attach a DUT to the test
points.
The primary display should show the value of the
DUT and the secondary display should show the %
difference from the standard.
6.
Repeat Step 5 for each DUT.
Note
: The range for the % difference is -99.9% to
99.9%. If the DUT falls outside that range, the
secondary display will show
OL
.
7.
To exit, press and hold the
REL%
key for
2 seconds.