PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT

 

Email: [email protected]

TEL:  (516) 334-5959 • FAX: (516) 334-5988

www.ietlabs.com

IET LABS, INC.

 HRRS 5kV AND 10kV SERIES

High Resistance, 5kV and 10kV

Decade Substituter

User and Service Manual

Copyright © 2017 IET Labs, Inc.

Visit www.ietlabs.com for manual revision updates

HRRS 5kV and 10kV im/Nov. 2017

Содержание HRRS 10kV Series

Страница 1: ...m TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC HRRS 5kV AND 10kV SERIES High Resistance 5kV and 10kV Decade Substituter User and Service Manual Copyright 2017 IET Labs Inc Visit www...

Страница 2: ...Email info ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC...

Страница 3: ...IET replaced at no charge when returned to IET Changes in this product not approved by IET or application of voltages or currents greater than those allowed by the specifications shall void this warra...

Страница 4: ...LTAGES 45 V ARE USED TAKE ALL MEASURES TO AVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARE CONDUCTORS WHEN USING THIS INSTRUMENT Use extreme cautio...

Страница 5: ...ainability and Reliability 6 4 2 Preventive Maintenance 6 4 3 Verification of Performance 6 4 3 1 Calibration Interval 6 4 3 2 General Considerations 6 4 3 3 Calibration Procedure 7 4 4 Schematic 7 4...

Страница 6: ...iv HRRS 5kV and 10kV Series This page is intentionally left blank...

Страница 7: ...m allowable applied input for each decade Custom requirements outside of the standard models can be satisfied To ensure low leakage the connection terminals are insulated with Kel F high resistance no...

Страница 8: ...0 03 0 10 70 700 70 700 5 10 100 k 1 M 0 01 0 03 0 10 230 2300 230 2300 5 10 1 M 10 M 0 01 0 03 0 10 1000 5000 1000 10 000 15 1 25 10 M 100 M 0 03 0 10 1 5000 5000 5000 10 000 10 000 25 1 50 High vol...

Страница 9: ...100 HRRS 6 1k 1 111 11 G 6 1 k HRRS 6 10k 11 111 1 G 6 10 k HRRS 6 100k 111 111 G 6 100 k HRRS 6 1M 1 111 11 T 6 1 M HRRS 6 10M 11 111 1 T 6 10 M Model Total resistance Number of decades Resolution H...

Страница 10: ...7 1 N9 68 67 787 5 23 5 7 1 8 CONSULT INSTRUCTION MANUAL FOR PROPER INSTRUMENT OPERATION WARNING HHS XQLW LQ D VHDOHG HQYLURQPHQW ZKHQ QRW LQ XVH 0DLQWDLQ ELQGLQJ SRVW DUHD FOHDQ IRU PLQLPXP HOHFWULFD...

Страница 11: ...an result in an irreversible change in resistance and require recalibration 3 1 Initial Inspection and Setup This instrument was carefully inspected before ship ment It should be in proper electrical...

Страница 12: ...y IET Labs or by a certified calibration laboratory If the user should choose to perform this procedure then the considerations below should be observed 4 3 2 General Considerations It is important wh...

Страница 13: ...ee Specifications on page 2 3 Confirm that the resistances fall within these limits If any resistances fall outside thee limits the associated switch assembly may require service or replacement WARNIN...

Страница 14: ...SW 10k 10 k step Decade Switch Assembly Not Shown HRRS SW 100k 100 k step Decade Switch Assembly Not Shown HRRS SW 1M 1 M step Decade Switch Assembly Not Shown HRRS SW 10M 10 M step Decade Switch Asse...

Страница 15: ......