 Maintenance

Chapter 4

MAINTENANCE

4.1  Maintainability and Reliability

It is possible to maintain models HRRS-5kV and 
HRRS-10kV inde

fi

 nitely.  They are reliable due to 

their closed design and sealed switches and resistors.  
The units are resistant to electromagnetic interference 
(EMI) because of their metal enclosure.

4.2  

Preventive Maintenance

Keep the units in a clean environment.  This will help 
prevent possible contamination.

The front panel should be periodically cleaned to 
eliminate any leakage paths from near or around the 
binding posts.  To clean the front panel:

Wipe the front panel clean using alcohol and a 
lint-free cloth.

4.3 Verifi cation of Performance

4.3.1 Calibration 

Interval

The HRRS 5kV and 10kV Series instruments should 
be veri

fi

 ed for performance at a calibration interval of 

twelve (12) months. This procedure may be carried 
out by the user if a calibration capability is available, 
by IET Labs, or by a certi

fi

 ed calibration laboratory. 

If the user should choose to perform this procedure, 
then the considerations below should be observed.

4.3.2 General 

Considerations

It is important, whenever testing the HRRS 5kV and 
10kV Series units, to be aware of the capabilities and 
limitations of the test instruments used. A resistance 
bridge may be employed, and there are direct-reading 
resistance meters or digital multimeters available 
that can verify the accuracy of these units, especially 
when used in conjunction with standards that can 
serve to con

fi

 rm or improve the accuracy of the test-

ing instrument

Such test instruments must have a 4:1 TUR  capabil-
ity for each value calibrated. Note that the accuracy 
varies for different decades. A number of commercial 
bridges and meters exist that can perform this task; 
consult IET Labs.

It is important to allow both the testing instrument 
and the HRRS-5kV/10kV Substituter to stabilize for 
a number of hours at the nominal operating tempera-
ture of 23°C, and at nominal laboratory conditions of 
humidity. There should be no temperature gradients 
across the unit under test.

Содержание HRRS 10kV Series

Страница 1: ...m TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC HRRS 5kV AND 10kV SERIES High Resistance 5kV and 10kV Decade Substituter User and Service Manual Copyright 2017 IET Labs Inc Visit www...

Страница 2: ...Email info ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC...

Страница 3: ...IET replaced at no charge when returned to IET Changes in this product not approved by IET or application of voltages or currents greater than those allowed by the specifications shall void this warra...

Страница 4: ...LTAGES 45 V ARE USED TAKE ALL MEASURES TO AVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARE CONDUCTORS WHEN USING THIS INSTRUMENT Use extreme cautio...

Страница 5: ...ainability and Reliability 6 4 2 Preventive Maintenance 6 4 3 Verification of Performance 6 4 3 1 Calibration Interval 6 4 3 2 General Considerations 6 4 3 3 Calibration Procedure 7 4 4 Schematic 7 4...

Страница 6: ...iv HRRS 5kV and 10kV Series This page is intentionally left blank...

Страница 7: ...m allowable applied input for each decade Custom requirements outside of the standard models can be satisfied To ensure low leakage the connection terminals are insulated with Kel F high resistance no...

Страница 8: ...0 03 0 10 70 700 70 700 5 10 100 k 1 M 0 01 0 03 0 10 230 2300 230 2300 5 10 1 M 10 M 0 01 0 03 0 10 1000 5000 1000 10 000 15 1 25 10 M 100 M 0 03 0 10 1 5000 5000 5000 10 000 10 000 25 1 50 High vol...

Страница 9: ...100 HRRS 6 1k 1 111 11 G 6 1 k HRRS 6 10k 11 111 1 G 6 10 k HRRS 6 100k 111 111 G 6 100 k HRRS 6 1M 1 111 11 T 6 1 M HRRS 6 10M 11 111 1 T 6 10 M Model Total resistance Number of decades Resolution H...

Страница 10: ...7 1 N9 68 67 787 5 23 5 7 1 8 CONSULT INSTRUCTION MANUAL FOR PROPER INSTRUMENT OPERATION WARNING HHS XQLW LQ D VHDOHG HQYLURQPHQW ZKHQ QRW LQ XVH 0DLQWDLQ ELQGLQJ SRVW DUHD FOHDQ IRU PLQLPXP HOHFWULFD...

Страница 11: ...an result in an irreversible change in resistance and require recalibration 3 1 Initial Inspection and Setup This instrument was carefully inspected before ship ment It should be in proper electrical...

Страница 12: ...y IET Labs or by a certified calibration laboratory If the user should choose to perform this procedure then the considerations below should be observed 4 3 2 General Considerations It is important wh...

Страница 13: ...ee Specifications on page 2 3 Confirm that the resistances fall within these limits If any resistances fall outside thee limits the associated switch assembly may require service or replacement WARNIN...

Страница 14: ...SW 10k 10 k step Decade Switch Assembly Not Shown HRRS SW 100k 100 k step Decade Switch Assembly Not Shown HRRS SW 1M 1 M step Decade Switch Assembly Not Shown HRRS SW 10M 10 M step Decade Switch Asse...

Страница 15: ......

Отзывы: