- 40 -
Inductance:
Range
Resolution
100/120Hz
1kHz
10kHz
100kHz
20.000
µ
H
0.001
µ
H
─
─
─
2.5%+5
200.00
µ
H
0.01
µ
H
─
─
1.2%+5
0.6%+3
2000.0
µ
H
0.1
µ
H
─
2.0%+5
0.6%+3
0.6%+3
20.000 mH
0.001 mH
1.2%+5
1.0%+5
0.3%+2
0.6%+3
200.00 mH
0.01 mH
0.3%+2
0.6%+3
0.3%+2
1.2%+5
2000.0 mH
0.1 mH
0.3%+2
0.3%+2
0.6%+3
─
20.000 H
0.001 H
0.3%+2
0.6%+3
1.2%+5
─
200.0 H
0.1 H
0.6%+3
1.2%+5
─
─
2.000 kH
0.001 kH
1.2%+5
─
─
─
*If reading <2000, unit on display is
µ
H
DCR:
Range
Resolution
Accuracy
200.00
Ω
0.01
Ω
1.0%+3
2.0000 k
Ω
0.0001 k
Ω
0.2%+2
20.000 k
Ω
0.001 k
Ω
0.2%+2
200.00 k
Ω
0.01 k
Ω
0.5%+2
2.0000 M
Ω
0.0001 M
Ω
1.0%+3
20.000 M
Ω
0.001 M
Ω
2.0%+3
200.0 M
Ω
0.1 M
Ω
2.0%+3
Accuracy v.s. Resistance (Z
DUT
):
DCR
100/120Hz
1kHz
10kHz
100kHz
0.1-1
Ω
1.2%+5
1.2%+5
1.2%+5
1.2%+5
2.5%+5
1-10
Ω
0.6%+3
0.6%+3
0.6%+3
0.6%+3
1.2%+5
10-100 k
Ω
0.3%+2
0.3%+2
0.3%+2
0.3%+2
0.6%+3
100 k-1 M
Ω
0.6%+3
0.6%+3
0.6%+3
0.6%+3
2.5%+5
1 M-20 M
Ω
1.2%+5
1.2%+5
1.2%+5
2.5%+5
2.5%+5
(2 M
Ω
max.)
>20 M
Ω
2.5%+5
2.5%+5
2.5%+5
─
─
Содержание DE-5000