Troubleshooting
104
Error message
Possible cause
Measures
Err 3
1. Blood pressure is beyond the
measurement range
1. Permanent notifications render
the blood pressure monitor
unsuitable for the patient.
2. Considerable arm movement
2. Keep the arm still during
measurement.
3. Problems with the pneumatics
3. If this error occurs continuously,
send the device to your specialist
supplier or directly to IEM GmbH
for checking.
Err 4
1. Data transmission cable not
correctly inserted into the blood
pressure monitor
1. Insert the cable correctly into the
blood pressure monitor (see
chapter 3.4 “Preparation for
measurement”).
2. Pins in the plug of the data
transmission cable are
mechanically damaged.
2. Check the plug to see whether
the pins on the inside are
damaged. If they are, call your
specialist supplier or IEM
3. Reading not transmitted
correctly
3. Start the transmission again.
Содержание Mobil-O-Graph PWA
Страница 2: ...Revision B 17 10 2018...
Страница 3: ...Gebrauchsanweisung DE Mobil O Graph PWA Mobil O Graph NG...
Страница 66: ...64...
Страница 67: ...Operating manual EN Mobil O Graph PWA Mobil O Graph NG...
Страница 129: ...Manual de instrucciones ES Mobil O Graph PWA Mobil O Graph NG...
Страница 191: ...Mode d emploi FR Mobil O Graph PWA Mobil O Graph NG...
Страница 254: ...252...
Страница 255: ...Istruzioni per l uso IT Mobil O Graph PWA Mobil O Graph NG...
Страница 320: ...318...
Страница 321: ...Gebruiksaanwijzing NL Mobil O Graph PWA Mobil O Graph NG...
Страница 352: ...Productbeschrijving 350 Geschatte levensduur van de manchet 6 maanden...
Страница 364: ...Probleemoplossing 362 Foutmelding Mogelijke oorzaak Maatregelen vergelijk deze met de manchetopdruk...
Страница 384: ......
Страница 385: ......