SECTION THREE
–
COMMANDS
IDRONAUT
–
Brugherio (MB)
OCEAN SEVEN 304
Plus
CTD
12-2015
29
Windows Personal Computer Preparation
The firmware is shipped together with the ITERM (IDRONAUT Terminal Emulation Programme).
The installation of ITERM is simple and does not require running any set-up.
Proceed as follows:
Copy the OS
304Plus
firmware upgrading package to a dedicated folder on the computer hard disk.
Unpack it using WINZIP or any other file uncompressing utility.
From the desktop and using the Windows resources, create a link on the desktop pointing to the
ITERM programme.
Run the ITERM programme. It starts with the correct communication parameters already set.
You are ready to switch ON the OS
304Plus
device.
Note
The ITERM port configuration is the following:
Communication speed
38400 bps
Data bit
8
Stop bit
1
Parity
None
Flow Control
None
The ITERM programme help contains the description of the programme main functions.
OS
304Plus
preparation
Connect the laboratory cable between the personal computer communication port and the OS
304Plus
and then switch on the OS304
Plus
. As soon as you switch the OS
304Plus
on, the start-up messages
must appear on the ITERM window.
Once the OS
304Plus
main menu appears, select the Service Menu and then select command 5. This
command allows the operator to access the “SETUP” using administrative rights. Once the command
is sent, the following message appears:
Access rights<<
Answer the request by typing “
ROOTACCESS
”. If the password is correctly entered and accepted
,
the menu appears again. From the shown service menu, select command 1
SETUP
! Go through the
parameters using the ENTER key until the request about the MAIN com port speed appears.
OCEAN SEVEN 304Plus - Setup
Configuration[191]-Status[177]-Calibration[309]{F8E5}
Warm-up timeout [s]:0 <
Number of rows between headers [1..255]:22 <
Main com. port BPS: 0)9k6,1)19k2,2)38k4,3)57k6,4)115k2:0 <