![idronaut Ocean Seven 304Plus Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/idronaut/ocean-seven-304plus/ocean-seven-304plus_operators-manual_618526017.webp)
SECTION ONE
–
SYSTEM DESCRIPTION
IDRONAUT
–
Brugherio (MB)
OCEAN SEVEN 304
Plus
CTD
12-2015
9
1.12.
SENSOR SPECIFICATION
The OS304
Plus
CTD can be equipped with the following sensors to measure:
Parameter
Range
Accuracy
Resolution
Time Constant
Pressure
0..1000 dbar
(1)
0.05 %F.S.
0.0015 % F.S.
50 ms
Temperature
-5.. +35 °C
0.002 °C
0.0001 °C
50 ms
+35.. +45 °C
0.01 °C
0.0001 °C
50 ms
(4)
Conductivity
0.. 90 mS/cm(*)
0.003 mS/cm
0.0003 mS/cm
50 ms
(2)
Analogue
0..5000 mV
1
mV
0.1
mV
50 ms
(3)
(1) Other standard pressure transducers available have : 10, 40, 100, 200, 500, 2000, 4000, 7000 dbar ranges.
(2)At 1 m/second flow rate.
(3) Six analogue inputs available for future expansion.
(4) OS304
Plus
–
MI version, special extended measurement range.
(*) By reducing the range to 0..70 mS/cm, the resolution becomes 0.0002 mS/cm.
1.13.
OPTIONAL SENSOR SPECIFICATIONS
The OCEAN SEVEN
304Plus
CTD can be optionally equipped with the IDRONAUT Highly Accurate
Precise (0.01%) pressure transducer, the IDRONAUT OEM Turbidity Meter and the IDRONAUT
dissolved oxygen sensor.
Parameter
Range
Accuracy
Resolution
Time Constant
Pressure
0..7000 dbar
0.01% F.S.
0.002 % F.S.
50 ms
Oxygen
0.. 50 pm
0.1 ppm
0.01 ppm
3s (*)
0.. 500 % sat.
1 % sat.
0.1 %sat.
3s (*)
Turbidity Meter
0.03..750 FTU/NTU
5 FTU/NTU
0.5 FTU/NTU
0.1 s
0.03..500 FTU/NTU
1 FTU/NTU
0.1 FTU/NTU
0.1 s
0.03..125 FTU/NTU
0.25 FTU/NTU
0.025 FTU/NTU
0.1 s
0.03.. 25 FTU/NTU
0.05 FTU/NTU
0.005 FTU/NTU
0.1 s
(*) from nitrogen to air.
A detailed description of the optional sensors can be found in the dedicated appendices.
1.14.
THE SENSORS
A short presentation of the standard IDRONAUT OCEAN SEVEN 304
Plus
sensors follows.
1.14.1.
The pressure sensor
The pressure sensor is a high quality strain gauge, centrally mounted on the CTD base, capable of
generating a linear signal output thus giving a resolution of 0.03% over the whole measuring range of
0 - 700 bar.
Type:
strain gauge.
Measurement range:
0..700 bar.
Accuracy:
0.05%FS.
Resolution
0.0015%FS.
Response time:
50 ms @1 m/s.
Measurement bridge resistance:
@ 25°C 3500
Ω
± 20%.
Excitation current:
0.6 mA.
Insulation:
@ 50 VCC 100
MΩ
Operating temperature:
-30..100 °C
Sensor body:
AISI 316 L.
Compensation:
Automatic compensation for temperature variations. Not
compensated for the barometric pressure variations.
Life:
unlimited.
Calibration:
offset calibration in air.
Содержание Ocean Seven 304Plus
Страница 1: ......