ECG Quality and Troubleshooting
46
Troubleshooting ECGs
Indistinct P-waves and movement artifact
P-waves may be “washed out” by artifacts caused by panting, shaking, and vocalizing.
Feline ECG with indistinct P-waves caused by purring
Canine ECG with respiratory (panting) artifact
To reduce movement artifact:
•
Move clips below the knees and elbows.
•
Distract the patient by tapping gently on the nose or offering a treat out of a closed hand.
•
Apply gentle pressure to the chest and/or legs.
•
Reposition the patient (if patient is standing, place in right-lateral position, or the reverse).
•
Record a chest lead to distinguish P-waves from movement artifacts.
Flat lines or indistinct complexes
Make sure clips have sufficient skin contact and alcohol saturation for conduction.
Flat lines and indistinct complexes