Olympus Cube Assembly Procedure 

 

 
 

 

 

11/11 

Step 

Action 

1.2.12

 

 

Place the filter inside the Exciter mounting area with the arrow pointing in.  

 

1.2.13

 

 

Insert the retaining ring open side into the Exciter mounting area of the cube. Using the 

Spanner Wrench, carefully rotate the retaining ring to avoid cross threading. If the ring will 

not turn, it is not threaded properly – slowly turn the ring out and try again. Do not force 

thread the ring as it will turn freely if oriented correctly. 

Tighten the ring snug against the Exciter filter. Once snug, tighten another quarter turn to 

secure. In the photo below, note the retaining ring shoulder side faces outward. 

 

Содержание Semrock U-MF2

Страница 1: ......

Страница 2: ...he Exciter side and one is on the Emitter side as shown below 1 1 2 Remove both filter retainer rings from the cube assembly 1 1 2 1 Each filter retaining ring has two slots where the spanner wrench b...

Страница 3: ...ed into the retaining ring slots 1 1 2 4 To remove the retaining ring insert the spanner wrench into the slots Applying even pressure across the ring carefully turn the ring counterclockwise to loosen...

Страница 4: ...e 1 2 Install the Semrock Filter Set 1 2 1 The diagram below illustrates a cross section of the Olympus cube for use in orienting and installing the Exciter Dichroic and Emitter filters into the cube...

Страница 5: ...arrow pointing inward The Dichroic reflective side is mounted facing the Exciter The Emitter is mounted with its arrow pointing outward Light passes through the cube in the direction of the arrows Not...

Страница 6: ...ure 6 11 Step Action 1 2 2 Remove the two Phillips Head screws to separate the cube halves Set the screws aside for reassembly later 1 2 3 Separate the cube halves to prepare for assembly of the Semro...

Страница 7: ...Note Do not remove the screws Loosen the screws to allow space between the housing and the clamp for installation of the Dichroic filter 1 2 5 Reflective Coated Side The filter edge is concealed by th...

Страница 8: ...n the mounting area and against the stop on the mounting clamp Alternately tighten the screws until they are snug Once snug slightly tighten the screws past their stopping point 1 2 7 Align and the cu...

Страница 9: ...Cube Assembly Procedure 9 11 Step Action 1 2 8 Orient the Emitter with the arrow pointing away from the cube as shown below 1 2 9 Place the filter inside the Emitter mounting area with the arrow poin...

Страница 10: ...g its position in the threads Keep the ring level across the threaded area while allowing the ring to tilt as it threads inward If the ring will not turn it is not threaded properly slowly turn the ri...

Страница 11: ...sing the Spanner Wrench carefully rotate the retaining ring to avoid cross threading If the ring will not turn it is not threaded properly slowly turn the ring out and try again Do not force thread th...

Страница 12: ...Olympus Cube Assembly Procedure 12 11 Step Action 1 2 14 Completed Olympus OMF Cube Assembly...

Отзывы: