12.
Go to the
“M
anual Dispense
”
option of a dispenser software, and dispense about 1ml from each
pump. Dispensing small amount is recommended after a long period of non-usage. Pay a good
attention during this process, and if error occurs, you may need to consult with your service contact.
Also, observer the pump nozzle area during the dispenses:
a.
To see if colorant is dispensed in an angle. If so, clean the nozzle area with a wet cloth / wet
wipe etc to remove dried colorants. Test again and ensure that the colorant is dispensed
straight down
b.
To see if colorant drops when the nozzle is opening or closing (as opposed to when the pump
is pushing the colorant out of nozzle). If so, it is possible that there is unwanted air in the
system. You may be able to eliminate the air by conducting large dispenser (approx. 200ml).
Test again and ensure no unwanted drop occurs during valve opening / closing
13.
Go to the “Maintenance” option, and conduct “Purge” from each pump. Pay a good attention during
this process, and if any error occurs, you may need to consult with your service contact.
14.
Check again if the brush is clean, and that there is correct amount of cleaning liquid (water / solvent)
in the brush container
Conduct a formula dispense as a test. Check if all peripherals are working (e.g. label printer,
spectrophotometer). If any issues, consult