Diode Testing
Function Resolution Accuracy Max. Test Max. Open
Current
Circuit
Voltage
1mV
+(1.5% +5)*
1.5mA
3V
* For 0.4V to 0.8V.
Overload Protection:
600V rms max
Diode Check:
To ensure a proper functioning diode, the meter will develop a
voltage across the component from a test current. The diode
test function allows measurements of forward voltage drops
across diode and transistor junctions.
1. Turn off power to the device or circuit that is being tested
and discharge all capacitors.
2. Plug the black test lead into the COM port and the red test
lead into the V
Ω
Hz port.
3. Set the rotary switch to the
Ω
position.
4. Press the blue button until is shown on the display
5. Connect the test leads to the diode. Normally the forward
voltage drop of a good silicone diode is shown between
400V and 90V. If the diode under test is defective, “000”
(short circuit) or “OL” (non-conductive) is displayed.
16
Capacitance
Range Resolution Max Display
Accuracy
4nF
1pF
4.000
±(3.% reading
40nF
10pF
40.00
+10 digits)
400nF
100pF
400.0
±(2% reading
4µF
1nF
4.000
+8 digits)
40µF
10nF
40.00
400µF
100nF
400.0
4mF
1µF
4.000
±(5% reading
40mF
10µF
40.00
+20 digits)
Overload Protection:
600V rms
To Measure Capacitance:
1. Plug the black test lead into the COM port and the red test
lead into the V
Ω
Hz port.
2. Set the rotary switch to the position.
3. Connect the test leads to the circuit to be measured.
4. Measure capacitance.
Hz
True RMS
15
Circuit Connection:
Hz
True RMS