12

Resistance (Ohms)

Range

Resolution

MaxDisplay

Accuracy

400

0.1

400.0

±(0.7% +3)

4k

1

4.000

±(0.4% +3)

40k

10

40.00

400k

100

400.0

4M

1k

4.000

±(0.6%+3)

40M

10k

40.00

±(1.5% +5)

Open Circuit Voltage: 

-1.3V approx.

To Measure Resistance:

1.  Turn the power off to the circuit or device that is to be 

measured and discharge all capacitors before attempting 
a measurement.

2.  Plug the black test lead into the COM port and the red test 

lead into the V

Hz port.

3.  Set the rotary switch to the 

position.

4.  For correct reading, ensure that the device being tested 

contains no voltage.

5.  Press the range button to select the proper range of the

meter.

6. Measure resistance.  

11

Frequency with      Hz Switch

Function

Range

Sensitivity

Accuracy

AC Voltage

400mV

40mV rms

0.01%

4V

0.2V rms

reading

40V

2V rms

+5 digits

400V

20V rms

750V

200V rms

AC Current

40mA

8mA rms

400mA

80mA rms

10A

8A rms

To Measure Frequency with the     Hz Switch

This switch is used to quickly view frequency during a mea-
surement of AC voltage or current.
1.  Push ~ Hz during an AC voltage or current measurement.
2.  The LCD will display the frequency measurement of the cir-

cuit.

3.  Push ~Hz again to return to the AC voltage or current mea-

surement.

Hz

True RMS

Hz

True RMS

Circuit Connection:

˜

˜

Содержание ND 2365-1

Страница 1: ...Hz True RMS 1 7 4 2 3 5 6 Symbols in this Manual This symbol indicates where cautionary or other information is found in the manual Fuse Battery Refer to Figure 1 and to the following numbered steps...

Страница 2: ...mode also changes the full scale range of the original voltage or cur rent mode WARNING 1 DO NOT UNDER ANY CIRCUMSTANCES EXCEED THESE RATINGS Voltage is not to exceed 1000 Volts Resistance Capacitance...

Страница 3: ...alarm 15 seconds before the meter automatically turns the power off To disable the auto power off function power up the meter while pressing either the range or Hz switch True RMS AC Volt Range Resolu...

Страница 4: ...test lead into the mA or A port 2 Set the rotary switch to the mA or A position 3 Push the blue button until AC is shown on the display 4 Connect the meter in series with the load or circuit 5 Measure...

Страница 5: ...load or circuit 5 Measure Frequency or RPM Overload Protection 600V rms Less than 50Hz the sensitivity is 1 5V DC Current Range Resolution Max Accuracy Voltage Display Burden 40mA 10 A 40 00 0 6 2 20...

Страница 6: ...the device being tested contains no voltage 5 Press the range button to select the proper range of the meter 6 Measure resistance 11 Frequency with Hz Switch Function Range Sensitivity Accuracy AC Vol...

Страница 7: ...00 Meter indicates actual resistance 3k Multiply meter display reading by 1 000 to acquire actual resistance 30k Multiply meter display reading by 1 000 to acquire actual resistance 300k Multiply mete...

Страница 8: ...ress the blue button until is shown on the display 5 Connect the test leads to the diode Normally the forward voltage drop of a good silicone diode is shown between 400V and 90V If the diode under tes...

Страница 9: ...Conditions Indoor Use Maximum Altitude 2000 Meter Installation Category IEC 1010 1000V Cat II 600V Cat III Pollution Degree 2 17 Continuity Check To Verify Continuity A continuity test ensures that al...

Страница 10: ...ent The 61 480 is powered by one 9V battery 1 Remove the test leads from the front terminals and turn the meter off 2 Remove the screw from the battery cover and lift to remove 3 Lift battery box and...

Отзывы: