Automatic Power Off

The thermometer will automatically shut off if left idle for more 
than 60 seconds unless it is in the PRB mode. If in the PRB 
mode, the unit will automatically shut off if left idle for more than 
12 minutes.

STORAGE 

 

The thermometer should be stored at room temperature between 
- 4 to 149°F (- 20 to 65°C).

LENS CLEANING

The lens should be kept clean at all times. Care should be 
taken when cleaning the lens using only a soft cloth or cotton 
swab with water. Allow the lens to fully dry before using the 
thermometer. 
Note: DO NOT submerge the unit in water.

LCD ERROR MESSAGES

The thermometer incorporates visual diagnostic messages as 
follows: ‘Hi’ or ’Lo’ is displayed when the temperature being 
measured is outside of the settings of HAL and LAL. ‘Er2’ is 
displayed when the thermometer is exposed to rapid changes 
in the ambient temperature. ‘Er3’ is displayed when the ambi-
ent temperature exceeds 32°F (0°C) or 122°F (+50°C). The 
thermometer should be allowed to stabilize to the working/
room temperature (minimum 30 minutes). 

Error 5~9

, for all 

other error messages it is necessary to reset the thermometer. To 
reset it, turn the instrument off, remove the battery and wait for 
a minimum of one minute, reinsert the battery and turn on. If the 
error message remains please contact the Service Department 
for further assistance.

BATTERIES

The thermometer incorporates visual low battery warning as 
follows:

  Batteries OK  

Batteries Low 

Batteries Dead

  measurements  

battery needs to be 

measurements are  

   are possible  replaced, measurements 

not possible

 

 

are still possible

 

 When the ‘Low Battery’ icon indicates the batteries are 

low, the batteries should be replaced immediately with two AAA, 

1.5V batteries. Please note: It is important to turn the instrument 

off before replacing the batteries otherwise the thermometer may 
malfunction.

7

SPECIFICATIONS

Item

Non-contact 

Infrared Scan 

function

Thermocouple Probe 

Scan function (K type; 

probe not included.)

Measurement  

Range

-76 to +932 °F 

(-60 to +500 °C)

-83.2 to +1999°F  

(-64 to +1400°C) 

Operating Range

32 to +122°F (0 to +50°C)

Accuracy 

[Tobj=59~95°F 

(15~35°C), 

Tamb=77°F 

(25°C)]

1.8°F (±1.0°C)

+/-1% of reading or 

1.8°F (1°C) whichever 

is greater [Test under

Tamb=73±10°F

(23±6°C)]

Accuracy 

[Tobj=-27~932° 

(-33~500°C), 

Tamb=73±5°F 

(23±3°C)]

±2% of reading 

or 4°F (2°C) 

whichever is 

greater

Emissivity Range

0.95 default – adjustable 0.1 to 1 step 

.01

Resolution 

14~392°F 

(-9.9~199.9°C)

0.1°F/0.1°C

Response Time 

(90%)

1sec

Distance: Spot

12:1(90% energy covered)

Battery Life

Typ.180, min 140 hours continuous use 

(Alkaline, without Laser and Back Light.)

Dimensions

48.8 x 132.7 x 146.0mm

(1.92×5.22×5.75 inch)

Weight

221.8 grams(7.8 oz) including batteries 

(AAA*2pcs)

Note: Under the electromagnetic field of 3V/m from 200 to 600 

MHz, the maximum error is 18°F (10°C).

  

N12966

5

Selecting a Function

Each time the mode key is pressed, the thermometer advances 
through the function cycle. 

Emissivity setting display (default is 0.95).

Emissivity adjustment - use the up and down 
keys to adjust emissivity to the desired value. 
Press the mode key to confirm the setting.

Maximum temperature reading is displayed.

Minimum temperature reading is displayed.

Difference between the maximum and 
minimum reading is displayed.

Average temperature is displayed.

High and low alarm setting - 
use the up and down keys to 
adjust the desired values. Press 
the trigger key to confirm the 
setting. When readings exceed 
the alarm values, the icon will 
flash and an audible beep will 
activate.

Connect the optional k-type thermocouple 
to the thermocouple socket. The thermometer 
will display the probe temperature auto-

matically. View the maximum and minimum 

probe measurements by pressing either the 
up or down keys.

Lock Mode

While in the E, MAX, MIN, dIF OR AVG functions, press the up 

key to turn on or off the lock mode. While in lock mode, the 
device will monitor temperatures for up to 60 minutes.

Changing to °F or °C

Press the down key to change the setting.

Backlit Function

While depressing the trigger key first, press the up key to turn 
the backlit on or off.

Laser Function

While depressing the trigger key first, press the down key to turn 
activate or de-activate the laser.

6

WARRANTY STATEMENT

This tester is warranted to the original purchaser against defects 
in material and workmanship for two years from the date of 
purchase. During this warranty period, IDEAL INDUSTRIES, INC. 
will, at its option, replace or repair the defective unit, subject to 
verification of the defect or malfunction. 
This warranty does not cover fuses, batteries or damage from 
abuse, neglect, accident, unauthorized repair, alteration, or 
unreasonable use of the instrument.
Any implied warranties arising out of the sale of an IDEAL 
product, including but not limited to implied warranties of 
merchantability and fitness for a particular purpose, are limited 
to the above. The manufacturer shall not be liable for loss of use 
of the instrument or other incidental or consequential damages, 
expenses, or economic loss, or for any claim or claims for such 
damage, expenses or economic loss.
State laws vary, so the above limitations or exclusions may not 
apply to you. This warranty gives you specific legal rights, and 
you may also have other rights which vary from state to state.
 

IDEAL INDUSTRIES, INC.

Sycamore, IL 60178, U.S.A.

877-201-9005 Technical Hotline 

www.idealindustries.com

ND 7055-1

 

Made in China