12
11
Resistance (Ohms)
Range
Resolution
MaxDisplay
Accuracy
400
Ω
0.1
Ω
400.0
±(0.7% +3)
4k
Ω
1
Ω
4.000
±(0.4% +3)
40k
Ω
10
Ω
40.00
400k
Ω
100
Ω
400.0
4M
Ω
1k
Ω
4.000
±(0.6%+3)
40M
Ω
10k
Ω
40.00
±(1.5% +5)
Open Circuit Voltage:
-1.3V approx.
To Measure Resistance:
1. Turn the power off to the circuit or device that is to be
measured and discharge all capacitors before attempting a
measurement.
2. Plug the black test lead into the COM port and the red test
lead into the °C°F V
Ω
Hz port.
3. Set the rotary switch to the
Ω
position.
4. For correct reading, ensure that the device being tested
contains no voltage.
5. Press the range button to select the proper range of the
meter.
6. Measure resistance.
True RMS
Circuit Connection:
Range Guide for Ohms (
Ω
):
400 = Meter indicates actual resistance
4k
= Multiply meter display reading by 1,000 to
acquire actual resistance.
40k
= Multiply meter display reading by 1,000 to
acquire actual resistance.
400k = Multiply meter display reading by 1,000 to
acquire actual resistance.
4M
= Multiply meter display reading by
1,000,000 to acquire actual resistance
400M = Multiply meter display reading by
1,000,000 to acquire actual resistance.
The meter displays total resistance through all possi-
ble paths between the probe-tips. These multiple
paths may result in measurements that do not corre-
spond to the ohm value indicated by the resistor
color code.
Determining Resistor Values:
To determine the value of a resistor, use the color bands on the
resistor and the table on the following page.