Frequency/RPM
Range
Resolution
Sensitivity
Accuracy
4.0KHz/40KRPM
1Hz/30RPM
100mV rms*
Frequency:
40KHz/400KRPM
10Hz/300RPM
0.01% ±1 digit
400KHz/4MRPM
100Hz/3KRPM
4MHz/40MRPM
1KHz/30KRPM
250mV rms
RPM:
40MHz/400MRPM
10KHz/300KRPM
IV rms
0.01% ±10 digits
Overload Protection: 600V rms
*Less than 20Hz the sensitivity is 1.5V
To Measure Frequency:
1. Plug the black test lead into the COM port and the red test
lead into the °C°F V
Ω
Hz port.
2. Set the rotary switch to the HZ RPM position.
3. For RPM, Push the blue button until RPM is shown on
the display.
4. Connect the meter in parallel with the load or circuit.
5. Measure Frequency or RPM.
10
9
DC Current
Range
Resolution
Max
Accuracy
Voltage
Display
Burden
40mA
10µA
40.00
±(0.6% +2) 200mVmax
400mA
0.1mA
400.0
±(0.7% +2)
2Vmax
10A
10mA
10.00
±(1.0% +3)
2V max
Overload Protection:
1A (500V) fast blow fuse for mA input
16A (500V) fast blow fuse for A input
To Measure DC Current:
1. Plug the black test lead into the COM port and the red test
lead into the mA or A port.
2. Set the rotary switch to the mA or A position.
3. Push the blue button until
DC
is shown on the display.
4. Connect the meter in series with the load or circuit.
5. Measure DC Current.
True RMS
True RMS
Circuit Connection:
Circuit Connection: