User Guide 

 

Page 28 

Cable Tests 

 

Wiremap 

 

 

Setup 

 

 

 

Cable Type  

– 

2 Pair 

 

 

 

 

 

- 4 Pair 568A 

 

 

 

 

 

- 4 Pair 568B 

 

 

 

Shield 

 

UTP 

 

 

 

 

 

- STP 

 

 

 

 

 

- UTP/STP 

 

 

 

 

 

 

 

NVP

 – Fixed 72% 

 

 

 

         - Custom 59% - 89% 

 

 

Termination Type 

 

 

 

None 

- Open  

 

 

 

 

Active Remote - 

#1 - #12 

 

 

Tests (No Termination) 

 

 

 

Faults

  

– Open circuit by pair 

 

 

 

  

- Short circuit by pin 

 

 

 

Length of pair 

 – Metres / Feet (Set in System Setup) 

 

 

 

 

 

- Range 3-100m / 10-330ft   

 

 

 

 

Tests (Active Remote Termination) 

 

 

 

I/D

 – Remote # 

 

 

 

Indications

 

on Remote

   – Voltage Warning  (>±10volts on any pins) 

 

 

 

 

 

- Pass/Fail 

 

 

 

Faults

  

– Open circuit by pin 

 

 

 

  

- Short circuit by pin 

 

 

 

 

- Crossed pairs 

 

 

 

 

- Split pairs 

 

 

 

 

- Bridged shorts 

 

 

 

 

- Remote shorts 

 

 

 

Length of pair 

 – Metres / Feet (Set in System Setup) 

 

 

 

 

 

- Range 3-100m / 10-330ft 

 

 

 

 

 

Tone Generator 

 

 

Setup 

 

 

 

No of

 

Tones – 3 

 

 

 

Wire I/D   – Tone applied to one of 8 pins relative to the other 7 

 

 

 

 

- Tone applied across one of 4 pairs 

 

 

Test 

 

 

 

Audible tone detected using compatible tone probe 

 

 

Содержание NaviTEK NT Plus

Страница 1: ...NaviTEK NT Plus Pro ...

Страница 2: ... be placed IDEAL INDUSTRIES Ltd pursues a policy of continual product improvement and reserves the right to alter without notice the specification design price or conditions of supply of any product or service iPhone and iTunes are trademarks of Apple Inc registered in the U S and other countries Google Play and Android are trademarks of Google Inc IDEAL INDUSTRIES LTD 2016 All rights reserved Pub...

Страница 3: ...e a live copper switch port or a live fiber switch port and runs tests appropriate to that configuration These tests are designed to give information about the port such as the switch MAC address and identification as well as to confirm that the port has been properly configured and is capable of reaching a number of strategic targets in the local network and the Internet The user may customize th...

Страница 4: ...it has reached the end of its life please contact your local IDEAL representative for service Power Module Recharging The power module can be fully recharged in three hours with the NaviTEK NT switched ON or OFF To recharge the power module connect the supplied power adaptor to the DC inlet For convenience the power module may be removed from or left attached to the unit for charging The Power LED...

Страница 5: ...it may be necessary to perform a master reset This will not delete any stored data 1 Remove the power module to access a small aperture in the NaviTEK NT 2 Insert a paper clip into the reset hole and press the internal reset switch 3 Replace the power module Replaceable insert RJ 45 socket To replace a damaged or worn RJ 45 socket insert proceed as follows Equipment required Kit IDEAL part number ...

Страница 6: ...le backup battery packs are also available Power button Press to switch the tester ON OFF Cable socket RJ45 socket to connect the tester to copper Ethernet cables and networks Ethernet LED Indicates that Ethernet is connected Activity LED Indicates Ethernet data activity Cursor buttons Press to change which screen function or item is highlighted Wi Fi Adapter Connects to smartphone app Charging In...

Страница 7: ...OBS to access the JOBS menu or following F3 MORE press F1 SETUP to access the SETUP menu see Menu Maps Auto Save Selection Highlight the icon using the cursor buttons and press ENTER to select how results should be saved MORE Press F3 MORE to access the SETUP menu using F1 SETUP or to control the tone generator using F2 TONE when a cable is connected Job Selection Shows the currently selected Job ...

Страница 8: ...phone or ISDN network the HOME screen displays an alarm and details of the voltages No further testing is possible until the voltages have been removed Indication of network type Indication of overall voltage Indication of pin voltages 1 2 3 4 5 6 7 8 Good Pair Indication Overall Cable Length Pair Lengths Indicates the lengths of the individual cable pairs Short Circuit Pair Indication Indicates a...

Страница 9: ...Remote ID number Overall Cable Length Result saved indication Indicates the name of the last saved result Wiremap test bar Indicates progress and final test result Green PASS Red FAIL Highlight the bar using the cursor keys then press ENTER to display the Wiremap screen Test Result Pair Lengths Indicates the lengths of the individual cable pairs Overall cable length FAULT Press F2 FAULT to display...

Страница 10: ...t result Green PASS Red FAIL Select then press ENTER to display the NET TEST screen in detail NET TEST Status Ready to test Test in progress Test Passed Test Failed Tester IP Status IP address assignment in progress Dynamic DHCP IP address assigned Static IP address assigned IP address assignment failed Select then press ENTER to display the IP screen Copper network connection TESTS Press F3 MORE ...

Страница 11: ...ils screen In the HOME screen select the IP icon then press ENTER to display the IP screen This screen shows detail of the IP status and address of the tester and the IP addresses of the network elements that are tested by the NET TEST Received Optical Power Select then press ENTER to display the Optical screen which gives details about the SFP fitted and the transmitted and received optical power...

Страница 12: ...ts The Secondary DNS is only tested if the Primary DNS Ping fails Select and press ENTER to display full details Gateway Ping result The Gateway is the route from the local network to the Internet Select and press ENTER to display full details Internet Ping result If this test passes the tested port has access to the Internet Select and press ENTER to display full details Trace Route result Select...

Страница 13: ... the tester to a mirror port or test port on the switch to monitor traffic in a particular network span MORE Press F1 MORE to display the next screen SCALE Press F2 SCALE to adjust the time scale MORE Press F1 MORE to display the next screen VLANs Lists the VLANs with ID detected in the network traffic VLAN ID The current VLAN setting of the tester Change VLAN Select a VLAN from the list of networ...

Страница 14: ...Guide Page 14 MORE Press F1 MORE to display the next screen Port data Shows information about the connection and the partner port MORE Press F1 MORE to display the next screen Error data Shows the number of Ethernet errors detected ...

Страница 15: ...d press ENTER to display the PoE screen Test result Test type Select PoE or PoE in the test setup screen Pair PoE may be presented on pairs 12 36 45 78 or both Test Data Shows the voltage current under load and power supplied by the port Test Status RUN Press F1 RUN to re run the PoE test without saving a result SETUP Press F3 SETUP to access the PoE test setup screen MORE Press F1 MORE to display...

Страница 16: ...he first Discovery Protocol message is received The screen then starts to flash the switch name and MAC address of the port that the Discovery Protocol message has come from If the message is confirmed as coming from the directly connected port the screen then shows full details of the port continuously At the end of 60 seconds from link establishment If a unique or confirmed Discovery Protocol me...

Страница 17: ...NaviTEK NT 151844 Iss 4 User Guide Page 17 Menu Maps ...

Страница 18: ...able Type o 2 4 Pairs o 568A 568B colour scheme Shield Type o UTP Shield must not be connected for test to pass o STP Shield must be connected for test to pass o UTP STP Test can pass if shield is connected or disconnected Custom NVP To achieve accurate length measurements set the NVP to match the cable type Set the details of the NET TEST Primary Secondary DNS and Gateway o Disabled The target is...

Страница 19: ... over Ethernet test Type o PoE Applies a load to draw current up to the maximum allowed for PoE o PoE Applies a load to draw current up to the maximum allowed for PoE o None PoE test disabled Min PoE power W o Enter the minimum power in watts for the PoE test to pass Min PoE power W o Enter the minimum power in watts for the PoE test to pass Set the details of the Ping 4 test Set the details of th...

Страница 20: ...orts contain both PDF and CSV files o PDF the reports contain only a PDF file o CSV the reports contain only a CSV file Size o Summary the reports contain only a summary table listing the overall result of each test o Brief the reports contain a summary table and a single page result for each test o Full the reports contain a summary table and full details of each test Results o All every test mad...

Страница 21: ...ry key containing an image named logo png with maximum size of 250 x 160 pixels Page 2 This is the Brief report for PORT0001 It shows that this port failed the Wiremap test Note the Job and Owner details Page 3 This is the Brief report for PORT0002 It shows that this port passed the Wiremap test Note the tester serial number Page 4 This is the Brief report for PORT0003 It shows that this port pass...

Страница 22: ...ts transfer Insert Wi Fi dongle into the NaviTEK NT USB port From the home screen press F1 JOBS The display will show the Job List screen Wi Fi connectivity is indicated by the top bar on the NaviTEK NT screen changing from grey to blue Now the NaviTEK NT is ready for results transfer wirelessly Note To minimise battery consumption the Wi Fi connectivity is only enabled for 5 minutes following pow...

Страница 23: ...TUP REPORTS screen You will be prompted for the NaviTEK NT Wi Fi password if it has been changed from the default value Once connected the App will display a list of JOBs on the NaviTEK NT These can be selected and downloaded to the smartphone Once results are on the iPhone they can then be transferred using email or other share mechanisms Apple is a trademark of Apple Inc registered in the U S an...

Страница 24: ...r type A USB type 1 1 Power Used for Battery charging Mains powering via adaptor Connector type 2 5mm pin power jack Polarity Centre pin positive Voltage 12v Current 2 A Location Bottom of optional power module Not present in standard alkaline battery pack Controls ON OFF Push button Used for Power ON OFF Function Keys F1 to F3 Used for Screen defined functions Navigation Keys Cursor and ENTER Use...

Страница 25: ...alkaline battery pack RJ45 Link LED Use ON indicates link UP Colour Green RJ45 Activity LED Use Flashing indicates link activity Colour Green Optical Link LED Use ON indicates Optical link UP Colour Green Optical Activity LED Use Flashing indicates Optical link activity Colour Green Ports RJ45 Setup Auto Negotiation Enabled Disabled Speed 10Mb s 100Mb s 1Gbps Mode Full Duplex Half Duplex MDI AUTO ...

Страница 26: ... Normal or Low Tests Ethernet Mode Ping4 Ping6 Trace Route4 Trace Route6 Hub Blink Netscan Loopback NET TEST Ping DNS Gateway Internet Trace Route Netscan Cable Mode Wiremap Tone Generator Auto Wiremap Service Detection Detected Services PoE 802 3af at Not Cisco pre standard ISDN S PBX Unknown Optical Supported SFPs The following SFP types are supported Use of other types of SFP is possible but co...

Страница 27: ...19 99 MAC Factory set VLAN Enabled Disabled VLAN ID 0 to 4094 VLAN Priority 0 to 7 802 1x Enabled Disabled EAP Method EAP MD5 EAP MSCHAPV2 EAP GTC EAP TLS EAP PEAP MD5 EAP PEAP MSCHAPV2 EAP PEAP GTC EAP PEAP TLS EAP TTLS MD5 EAP TTLS MSCHAPV2 EAP TTL GTC EAP TTLS TLS Username Password Certificate Import password Root CA certificate Tests Optical Tx Power dBm using a specified SFP Rx Power dBm usin...

Страница 28: ...in System Setup Range 3 100m 10 330ft Tests Active Remote Termination I D Remote Indications on Remote Voltage Warning 10volts on any pins Pass Fail Faults Open circuit by pin Short circuit by pin Crossed pairs Split pairs Bridged shorts Remote shorts Length of pair Metres Feet Set in System Setup Range 3 100m 10 330ft Tone Generator Setup No of Tones 3 Wire I D Tone applied to one of 8 pins relat...

Страница 29: ...cal 128bit HEX IP address Network Prefix 64 bit 128 bit Pingv4 Setup Target Numerical address URL Store up to 10 Count 1 to 999999 Pause 1 to 5 Sec Length 8 to 1000 bytes Results Info READY IN PROGRESS PASSED NO RESPONSE UNKNOWN HOST Tx Count 1 to 999999 Rx Count 1 to 999999 Delay ms Minimum Average Maximum Pingv6 Setup Target IPv6 address URL Store up to 10 Count 1 to 999999 Pause 1 to 5 Sec Leng...

Страница 30: ...ximum Trace Routev4 Setup Target Numerical address URL Max Hops 2 to 100 Timeout 2 to 30 sec Type ICMP UDP Results Info READY IN PROGRESS PASSED NO RESPONSE UNKNOWN HOST Hop Numerical address Delay ms t1 t2 t3 Trace Routev6 Setup Target Numerical address URL Max Hops 2 to 100 Timeout 2 to 30 sec Type UDP Results Info READY IN PROGRESS PASSED NO RESPONSE UNKNOWN HOST Hop Numerical address Delay ms ...

Страница 31: ... Sequence Off 10 Off 100 Off 1000 Mb s RJ 45 Off On Optical Loop Setup Loop Type Wireline MAC IP UDP All Traffic Yes No Statistics IP Results IPv4 info listening assigned DHCP failed DHCP or Static IPv4 Address IPv4 Netmask IPv4 Gateway IPv4 DNS1 IPv4 DNS2 IPv6 Enabled or Disabled info listening assigned DHCP failed Stateful DHCPv6 or Stateless or Static IPv6 Address IPv6 Network Prefix 64 bit or ...

Страница 32: ...vel VLAN ID Rx 802 1x Status Auth Not Started Auth Started Auth Completed Successfully Auth Failed Connected Successfully auth Port Status Unauthorised Authorised EAP Method Used Key Management Used LINK Results PORT PoE Voltage 0 60V PoE Pairs 12 36 or 45 78 Speed Duplex MDI MDIX Signal Level Polarity PARTNER 10M HD 10M FD 100M HD 100M FD 1000M HD 1000M FD continued ...

Страница 33: ...rage Configurations Internal storage Number of configurations 2 Current Factory settings Export Import Port USB Format xml Certificates 802 1x Max number 10 Results Internal storage Max Number of Jobs Projects 50 Max Number of result sets per Job 5000 depending on tests performed Max total number of result sets Up to 5000 depending on tests performed Export Port USB Wi Fi Format PDF CSV summary on...

Страница 34: ...ormat 12 hour 24 hour Software update Upgrade Via USB General Date Time Internal Clock Used for Timestamping results Autonomy Up to 1 day with battery removed Power Battery Supported Types Standard power module 4 x AA NiMH cells Alkaline battery pack with 4 AA cells Autonomy Up to 5 hours power module only Recharge time 3 hours Power module only Battery level Indication Full 2 3 1 3 Empty Physical...

Страница 35: ...ral continued Environmental Temperature Operating 0 C to 40 C Storage 20 C to 70 C Relative Humidity Min 5 Max 90 non condensing Approvals EMC EN 55022 2006 A1 2007 EN55024 1998 A1 2001 A2 2003 Safety IEC 60950 1 2005 A1 2009 EN 60950 1 2006 A1 2010 ...

Страница 36: ...Pv6 Stateless P address assigned automatically using Stateless Address Autoconfiguration SLAAC without DHCPv6 Static IP address assigned manually by the operator LAN Local Area Network MAC Media Access Control MDI Medium Dependent Interface MDIX Medium Dependent Interface Crossover NVP Nominal Velocity of Propagation of signals in a cable expressed as a percentage of the speed of light in a vacuum...

Страница 37: ...NaviTEK NT 151844 Iss 4 User Guide Page 37 Term Description Wi Fi Wireless Network ...

Страница 38: ......

Отзывы: